Falsely diagnosed thyrotoxicosis caused by anti-streptavidin antibodies and pre-wash procedures.

Ishikawa Takuya, Sakai Hiroyuki, Itaya Tokutaro, Hirotsugu Suwanai, Shikuma Jumpei, Miwa Takashi, Suzuki Ryo, Odawara Masato

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Anti-streptavidin antibodies are causal determinants of analytical interference during Thyroid function tests, and numerous reports have detailed such interference, with anti-streptavidin antibodies attracting attention.

CASE PRESENTATION: We conducted a straightforward investigation of interference due to anti-streptavidin antibodies, with a case of a 60-year-old Japanese man who consulted our department for inconsistencies between his clinical course and Thyroid function tests. Experiments were conducted using Cobas8000 e602, which employs assay procedures with pre-wash to evaluate FT4 and FT3 levels.

CONCLUSIONS

To our knowledge, this is the first published report to clearly investigate such interferences using a combination of polyethylene glycol precipitation, heterophilic blocking tube precipitation, streptavidin-coated magnetic particle precipitation, and different instruments with or without pre-wash. Clinicians should consider that interferences caused by anti-streptavidin antibodies could lead to a misdiagnosis of thyrotoxicosis. Moreover, discussions between laboratory specialists, clinicians, and manufacturers are required to identify interferences and avoid unnecessary examinations and inappropriate treatment.

Для цитирования: Ishikawa Takuya, Sakai Hiroyuki, Itaya Tokutaro, Hirotsugu Suwanai, Shikuma Jumpei, Miwa Takashi, Suzuki Ryo, Odawara Masato. Falsely diagnosed thyrotoxicosis caused by anti-streptavidin antibodies and pre-wash procedures.. Thyroid research. 2021. DOI: 10.1186/s13044-021-00108-y.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.