DOI 10.1016/j.cca.2026.121287

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2026

Spurious undetectable albumin due to hemodialysis catheter lock contamination.

Namlı Ahmet, Turgut Hasan Alp, Şeneş Mehmet

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Contamination of blood drawn through a hemodialysis catheter with heparin lock solution can interfere with albumin measurement by the bromocresol green (BCG) method, which shows dose-dependent negative interference with heparin. Catheter lock solutions (1000-5000 U/mL) are far more concentrated than the heparin used in standard collection tubes and pose a correspondingly greater risk of interference.

CASE: In a 46-year-old man on maintenance hemodialysis, a routine sample drawn from a tunneled permanent catheter without discard of the lock solution yielded an albumin below the analytical measuring range (<2 g/L) in a clinically stable patient with no signs of severe hypoalbuminemia. Analytical error was excluded by an unchanged repeat measurement and acceptable internal quality control performance. A repeat sample, drawn after an appropriate discard volume, gave an albumin of 42 g/L, confirming heparin contamination as the cause.

CONCLUSION

Catheter lock solution heparin can produce an analytically uninterpretable albumin result by the BCG method. Adherence to evidence-based discard volume protocols is essential in dialysis units.

Для цитирования: Namlı Ahmet, Turgut Hasan Alp, Şeneş Mehmet. Spurious undetectable albumin due to hemodialysis catheter lock contamination.. Clinica chimica acta; international journal of clinical chemistry. 2026. DOI: 10.1016/j.cca.2026.121287.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.