EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2021
Near-infrared spectroscopy carotid plaque characteristics and cerebral embolism in carotid artery stenting.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Nakagawa Ichiro, Kotsugi Masashi, Park Hun Soo, Furuta Takanori, Sato Fumiya, Myochin Kaoru, Nishimura Fumihiko, Yamada Syuichi, Motoyama Yasushi, Nakase Hiroyuki
Аннотация
BACKGROUND: Perioperative thromboembolism is the main consideration in carotid artery stenting (CAS). Precise evaluation of carotid plaque components is clinically important to reduce ischaemic complications since CAS mechanically pushes plaque outwards, which releases plaque debris into the bloodstream.
AIMS: This study aimed to determine whether high lipid core plaque (LCP) assessed by catheter-based near-infrared spectroscopy (NIRS) is associated with ipsilateral cerebral embolism by diffusion-weighted magnetic resonance imaging during CAS using a first-generation stent.
METHODS: Carotid stenosis magnetic resonance (MR) T1-weighted plaque signal intensity ratio (T1W-SIR) followed by NIRS assessment at the time of CAS (using the carotid artery Wallstent) was performed in 117 consecutive patients.
RESULTS: The maximum lipid core burden index (max-LCBI) at minimal luminal areas (MLA; max-LCBIMLA) and the max-LCBI for any 4 mm segment in a target lesion defined as max-LCBIarea were significantly higher for the post-procedural new ipsilateral diffusion-weighted magnetic resonance imaging (DWI)-positive than negative patients (p<0.001 for all). There was a significant linear correlation between max-LCBIarea and the number of new emboli (r=0.544, p<0.0001). We also found that the second quantile (Q2) of T1W-SIRMLA had a significantly higher max-LCBIMLA and a higher incidence of DWI positivity than Q1 and Q3 (p<0.001 for all). Furthermore, max-LCBIMLA appeared to distinguish between patients with and without postoperative new ipsilateral DWI positivity (AUC 0.91, 95% CI: 0.86-0.96; p<0.0001).
CONCLUSIONS: High LCP assessed by NIRS is associated with cerebral embolism by diffusion-weighted imaging in CAS using a first-generation stent.
AIMS: This study aimed to determine whether high lipid core plaque (LCP) assessed by catheter-based near-infrared spectroscopy (NIRS) is associated with ipsilateral cerebral embolism by diffusion-weighted magnetic resonance imaging during CAS using a first-generation stent.
METHODS: Carotid stenosis magnetic resonance (MR) T1-weighted plaque signal intensity ratio (T1W-SIR) followed by NIRS assessment at the time of CAS (using the carotid artery Wallstent) was performed in 117 consecutive patients.
RESULTS: The maximum lipid core burden index (max-LCBI) at minimal luminal areas (MLA; max-LCBIMLA) and the max-LCBI for any 4 mm segment in a target lesion defined as max-LCBIarea were significantly higher for the post-procedural new ipsilateral diffusion-weighted magnetic resonance imaging (DWI)-positive than negative patients (p<0.001 for all). There was a significant linear correlation between max-LCBIarea and the number of new emboli (r=0.544, p<0.0001). We also found that the second quantile (Q2) of T1W-SIRMLA had a significantly higher max-LCBIMLA and a higher incidence of DWI positivity than Q1 and Q3 (p<0.001 for all). Furthermore, max-LCBIMLA appeared to distinguish between patients with and without postoperative new ipsilateral DWI positivity (AUC 0.91, 95% CI: 0.86-0.96; p<0.0001).
CONCLUSIONS: High LCP assessed by NIRS is associated with cerebral embolism by diffusion-weighted imaging in CAS using a first-generation stent.