Acute hypernatraemia during bicarbonate-buffered haemodialysis.

Williams D J, Jugurnauth J, Harding K, Woolfson R G, Mansell M A

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Five patients on maintenance haemodialysis were exposed to varying degrees of hypernatric dialysate, leading to acute hypernatraemia (plasma sodium concentrations 158 mmol/l to 179 mmol/l). With the exception of one patient, who developed pulmonary oedema, symptoms were minimal and in each case hypernatraemia was corrected without residual complications. The hypernatric dialysate resulted from a granular and less soluble batch of sodium bicarbonate powder. The extra effort required to dissolve the powder caused CO2 to be shaken out of solution, producing sodium carbonate and raising the pH. Mixing calcium from the 'acid' concentrate with excess carbonate in the 'bicarbonate' concentrate led to rapid precipitation of calcium carbonate on the conductivity monitoring cells. Dialysate conductivity was incorrectly sensed as low by the coated conductivity cells, so that an increasing amount of 'acid' concentrate, with its accompanying electrolytes, was delivered to the patient. When the granular powder was ground to a fine powder, passed through a 125 microns sieve and gently dissolved, the machine operated normally. We recommend that sodium bicarbonate powder is supplied with a sieve size no greater than 125 microns, kept dry to prevent the formation of large crystals, and dissolved gently.

Для цитирования: Williams D J, Jugurnauth J, Harding K, Woolfson R G, Mansell M A. Acute hypernatraemia during bicarbonate-buffered haemodialysis.. Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association. 1994. DOI: 10.1093/ndt/9.8.1170.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.