Validation of rotational thromboelastometry during cardiopulmonary bypass: A prospective, observational in-vivo study.

Gronchi Fabrizio, Perret Anouk, Ferrari Enrico, Marcucci Carine M, Flèche Jérome, Crosset Monique, Schoettker Patrick, Marcucci Carlo

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Rotational thromboelastometry (ROTEM) is a whole blood point-of-test used to assess the patient's coagulation status. Three of the available ROTEM tests are EXTEM, INTEM and HEPTEM. In the latter, heparinase added to the INTEM reagent inactivates heparin to reveal residual heparin effect. Performing ROTEM analysis during cardiopulmonary bypass (CPB) might allow the anaesthesiologist to anticipate the need for blood products.

OBJECTIVE

The goal of this study was to validate ROTEM analysis in the presence of very high heparin concentrations during CPB.

DESIGN

Prospective, observational trial.

SETTING

Single University Hospital.

PARTICIPANTS

Twenty patients undergoing coronary artery bypass grafting.

MAIN OUTCOME MEASURE: ROTEM analysis was performed before heparin administration (T0), 10 min after heparin (T1), at the end of CPB (T2) and 10 min after protamine (T3). The following tests were performed: EXTEM, INTEM, and HEPTEM. Heparin concentrations were measured at T1 and at the end of bypass (T2).

RESULTS

At T1, EXTEM differed from baseline for coagulation time: +26.7 s (18.4 to 34.9, P < 0.0001), α: -3° (1.0 to 5.4, P = 0.006) and A10: -4.4 mm (2.3 to 6.5, P = 0.0004). INTEM at T0 was different from HEPTEM at T1 for coagulation time: + 47 s (34.3 to 59.6, P > 0.0001), A10: -2.3 mm (0.5 to 4.0, P = 0.01) and α -2° (1.0 to 3.0; P = 0.0007). At T2, all parameters in EXTEM and HEPTEM related to fibrin-platelet interaction deteriorated significantly compared to T1. At T3, EXTEM and INTEM were comparable to EXTEM and HEPTEM at T2.

CONCLUSION

HEPTEM and EXTEM measurements are valid in the presence of very high heparin concentrations and can be performed before protamine administration in patients undergoing cardiac surgery with CPB.

TRIAL REGISTRATION: clinicaltrials.gov Identifier: NCT01455454.

Для цитирования: Gronchi Fabrizio, Perret Anouk, Ferrari Enrico, Marcucci Carine M, Flèche Jérome, Crosset Monique, Schoettker Patrick, Marcucci Carlo. Validation of rotational thromboelastometry during cardiopulmonary bypass: A prospective, observational in-vivo study.. European journal of anaesthesiology. 2014. DOI: 10.1097/eja.0b013e328363171a.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.