DOI 10.1016/j.jaccas.2024.103197
Buckling Under Pressure: Evolut FX Delivery System Malfunction and Failure to Recapture During TAVR.
Аннотация
OBJECTIVE
We present a case of transcatheter aortic valve replacement in a 70-year-old woman with severe bicuspid aortic valve stenosis complicated by incomplete recapture of the Evolut valve and buckling of the delivery system.
KEY STEPS: Our key steps included the following: 1) unsuccessful withdrawal of the buckled delivery system and valve into the inline sheath; 2) transradial snaring of the delivery system, followed by traction from both ends to reduce the amount of buckling; and 3) once elongated as much as possible, the malfunctioning Evolut delivery system and valve were removed as a unit over the wire.
POTENTIAL PITFALLS: Aortic, iliac, or femoral injury: Reliant balloon was placed in the distal aorta use in case of arterial rupture.
TAKE-HOME MESSAGES: The Evolut delivery system can buckle and accordion under tension in complex aortic anatomy. A technique of snaring from above and tension from below can mitigate the buckled area and allow removal of the damaged delivery system via the endovascular route.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.