Invention of a new percutaneous closure technique for vascular haemostasis in percutaneous veno-arterial extracorporeal membrane oxygenation.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Guanqiang Li, Bo Hu, Yuan Sun, Ming Chen, Xicheng Zhang, Xianchen Huang
Аннотация
OBJECTIVE: To assess the feasibility and safety of a novel closure technique, namely the parallel closure technique (Par-CT), in which the sheath is placed parallel to the ProGlide system during veno-arterial extracorporeal membrane oxygenation (VA-ECMO) to enhance the efficiency of percutaneous vascular haemostasis.
METHODS: The abdominal aortas of goats were punctured with a 17F sheath to simulate arterial cannula vascular access for VA-ECMO, and the Par-CT was used to achieve vascular haemostasis. In five goats, the abdominal aorta was punctured with a 17F sheath and then a 6 to 10F vascular sheath was introduced. This resulted in a smaller vascular rupture area and successful closure of the aortic rupture using the Par-CT.
RESULTS: If the sheath tube that was introduced was < 8 F, the puncture site could not be closed and would therefore have to be resutured and the Par-CT would have to be repeated. However, the puncture site could be sutured appropriately if the sheath tube that was introduced was ≥ 8F.
CONCLUSIONS: This study suggested that the Par-CT is both safe and effective for haemostasis after weaning from VA-ECMO.
METHODS: The abdominal aortas of goats were punctured with a 17F sheath to simulate arterial cannula vascular access for VA-ECMO, and the Par-CT was used to achieve vascular haemostasis. In five goats, the abdominal aorta was punctured with a 17F sheath and then a 6 to 10F vascular sheath was introduced. This resulted in a smaller vascular rupture area and successful closure of the aortic rupture using the Par-CT.
RESULTS: If the sheath tube that was introduced was < 8 F, the puncture site could not be closed and would therefore have to be resutured and the Par-CT would have to be repeated. However, the puncture site could be sutured appropriately if the sheath tube that was introduced was ≥ 8F.
CONCLUSIONS: This study suggested that the Par-CT is both safe and effective for haemostasis after weaning from VA-ECMO.