OBJECTIVES

In contrast to stented transcatheter aortic valves, the Direct Flow Medical (DFM) valve is a stentless bovine aortic bioprosthesis mounted in a non-metallic inflatable frame. Hence, severe asymmetric annular calcification may result in residually elevated transaortic pressure gradients after DFM implantation. We present a novel intraprocedural dilatation (IDIL) technique for successful implantation of the DFM valve in the presence of complex annular calcification.

METHODS

Between January 2014 and May 2015, 55 patients underwent DFM valve-based transcatheter aortic valve implantation at our institution. Of these, 5 patients required an IDIL technique due to a residual intraoperative transaortic pressure mean gradient above 15 mmHg. The mean patient age was 73 ± 8.2 years; the mean logistic EuroSCORE was 24.5 ± 8.2% and the mean Society of Thoracic Surgeons score was 6.3 ± 4.3%.

RESULTS

The IDIL technique immediately attenuated transvalvular mean pressure gradients from 20 ± 2 mmHg to 6 ± 1 mmHg. The results remained stable during the 30-day observation period at 10 ± 3 mmHg. Minimal paravalvular aortic regurgitation (trace) was detected in 2 patients. No in-hospital deaths were observed.

CONCLUSIONS

The IDIL technique facilitates safe DFM valve implantation in patients with complex asymmetric annular calcification without adverse side effects on valve structure or performance in short-term follow-up.

Для цитирования: Bushnaq Hasan, Raspé Christoph, Öner Alper, Yücel Seyrani, Ince Hüseyin, Sommer Sebastian-Patrick. A new technique to implant a transcatheter inflatable, fully repositionable prosthesis in aortic stenosis with severe asymmetric calcification.. Interactive cardiovascular and thoracic surgery. 2017. DOI: 10.1093/icvts/ivx197.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.