[Are we measuring the right intraocular pressure after excimer laser photorefractive laser keratoplasty in myopia?].

Schipper I, Senn P, Niesen U

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

It is possible that the traditional method to determine the intraocular pressure after excimer-laser PRK is inaccurate. Measuring the pressure in the temporal part of the cornea might give the true values.

METHODS

Intraocular pressure was measured with a Goldmann Applanation Tonometer and with the Tonopen, before and after PRK for myopia in the central and in the temporal parts of the cornea. The paired student t-test was used for statistical analysis.

RESULTS

The results of central and temporal measurements before treatment were identical with both instruments. After PRK, central values were 2 to 3 mm Hg lower than temporal values when measured with a Goldmann Tonometer, and about 2 mm lower when measured with the Tonopen. The differences were highly significant (p < 0.0001 and p = 0.004 respectively).

CONCLUSIONS

The intraocular pressure measured in the usual manner after excimer-laser PRK is lower than the temporally measured pressure. These differences could be caused by absence of the Bowman's membrane, thinning of the cornea and/or change of its topography.

Для цитирования: Schipper I, Senn P, Niesen U. [Are we measuring the right intraocular pressure after excimer laser photorefractive laser keratoplasty in myopia?].. Klinische Monatsblatter fur Augenheilkunde. 1995. DOI: 10.1055/s-2008-1035452.