DOI 10.1080/14740338.2018.1442431
The nature, magnitude, and reporting compliance of device-related events for intravenous patient-controlled analgesia in the FDA Manufacturer and User Facility Device Experience (MAUDE) database.
Аннотация
BACKGROUND
The aim of this study is to determine the characteristics, magnitude, and the quality of reporting of mandated events involving intravenous patient-controlled analgesia (IV-PCA) devices in the Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database; a postmarket surveillance system.
METHODS
We utilized a mixed-methods approach to systematically characterize structured data and text narratives associated with IV-PCA events submitted to MAUDE between 1 January 2011 and 12 September 2016.
RESULTS
Of 1,430 IV-PCA events reported during the study period, 6.4% were adverse events (AEs) as identified via structured data fields in the MEDWATCH forms. Upon qualitative review of the narrative texts, 11.0% of events were associated with an unfavorable clinical outcome, which was 71% higher than the incidence of the adverse outcomes reported using the structured data fields. Device-related issues, which were mostly preventable, accounted for 86.9% of events. Of 65 reportable events submitted by manufacturers, 18.5% did not comply with reporting requirements as mandated by law.
CONCLUSION
Patients on IV-PCA continue to experience serious complications as a result of preventable errors. Multi-modal interventions including educational training and the development and adoption of PCA devices with improved safety features are needed to improve safety.