MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-2008-1038722

Медицинская библиотека · The Thoracic and cardiovascular surgeon · 2008

A staged approach towards interventional aortic valve implantation with a sutureless valve: initial human implants.

Shrestha M, Khaladj N, Bara C, Hoeffler K, Hagl C, Haverich A

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Percutaneous implantable aortic valves may become an alternative to conventional approaches. The purpose of this study was to assess a new sutureless aortic valve (Perceval Sorin). As a first step, an open approach using cardiopulmonary bypass (CPB) was chosen to evaluate the feasibility of implantation.

METHODS

Between April and September 2007, 16 high-risk patients (13 females, aged 81 [76 - 88]) were operated on via a median sternotomy, using CPB and cardioplegia (Euro Score 17 [8 - 73]). All patients had significant aortic valve disease and seven of these patients had concomitant coronary artery disease. This pilot project was initiated with prior approval of the Institutional Review Board. All patients gave informed consent.

RESULTS

One patient died during hospital stay for unknown reasons. Autopsy revealed no valve related pathologies. CPB time was 60 min (41 - 130), cross-clamping time was 36 (22 - 79) min. Intraoperative as well as postoperative echocardiography revealed neither aortic insufficiency nor paravalvular leakage in any of the patients.

CONCLUSIONS

The new approach as described here is a technically simple alternative to conventional aortic valve replacement in high-risk patients and offers the potential of less invasive approaches. It appears especially useful in patients with severe calcification of the aortic root. CPB and cross-clamping times were markedly reduced compared with patients who underwent conventional operations.

Для цитирования: Shrestha M, Khaladj N, Bara C, Hoeffler K, Hagl C, Haverich A. A staged approach towards interventional aortic valve implantation with a sutureless valve: initial human implants.. The Thoracic and cardiovascular surgeon. 2008. DOI: 10.1055/s-2008-1038722.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.