DOI 10.1371/journal.pone.0221965
Comparison of hemodynamic effects and resuscitation outcomes between automatic simultaneous sterno-thoracic cardiopulmonary resuscitation device and LUCAS in a swine model of cardiac arrest.
Аннотация
INTRODUCTION
Mechanical cardiopulmonary resuscitation (CPR) devices are widely used to rescue patients from cardiac arrest. This study aimed to compare hemodynamic effects and resuscitation outcomes between a motor-driven, automatic simultaneous sterno-thoracic cardiopulmonary resuscitation device and the Lund University cardiac arrest system (LUCAS).
MATERIAL AND
METHODS
After 2 minutes of electrically induced ventricular fibrillation (VF), Yorkshire pigs (weight 35-60 kg) received CPR with an automatic simultaneous sterno-thoracic CPR device (X-CPR group, n = 13) or the Lund University cardiac arrest system (LUCAS group, n = 12). Basic life support for 6 minutes and advanced cardiovascular life support for 12 minutes, including defibrillation and epinephrine administration, were provided. Hemodynamic parameters and resuscitation outcomes, including return of spontaneous circulation (ROSC), 24-hour survival, and cerebral performance category (CPC) at 24 hours, were evaluated.
RESULTS
Hemodynamic parameters, including aortic pressures, coronary perfusion pressure, carotid blood flow, and end-tidal carbon dioxide pressure were not significantly different between the two groups. Resuscitation outcomes were also not significantly different between the groups (X-CPR vs. LUCAS; rate of ROSC: 31% vs 25%, p = 1.000; 24-hour survival rate: 31% vs 17%, p = 0.645; neurological outcome with CPC ≤2: 31% vs 17%, p = 0.645). Also no significant difference in incidence complications associated with resuscitation was found between the groups.
CONCLUSIONS
CPR with a motor-driven X-CPR and CPR with the LUCAS produced similar hemodynamic effects and resuscitation outcomes in a swine model of cardiac arrest.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.