Медицинская библиотека · American journal of critical care : an official publication, American Association of Critical-Care Nurses · 2019

Contribution of Electrocardiographic Accelerated Ventricular Rhythm Alarms to Alarm Fatigue.

Suba Sukardi, Sandoval Cass Piper, Zègre-Hemsey Jessica K, Hu Xiao, Pelter Michele M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Excessive electrocardiographic alarms contribute to "alarm fatigue," which can lead to patient harm. In a prior study, one-third of audible electrocardiographic alarms were for accelerated ventricular rhythm (AVR), and most of these alarms were false. It is uncertain whether true AVR alarms are clinically relevant.

OBJECTIVES

To determine from bedside electrocardiographic monitoring data (1) how often true AVR alarms are acknowledged by clinicians, (2) whether such alarms are actionable, and (3) whether such alarms are associated with adverse outcomes ("code blue," death).

METHODS

Secondary analysis using data from a study conducted in an academic medical center involving 5 adult intensive care units with 77 beds. Electronic health records of 23 patients with 223 true alarms for AVR were examined.

RESULTS

The mean age of the patients was 62.9 years, and 61% were white and male. All 223 of the true alarms were configured at the warning level (ie, 2 continuous beeps), and 215 (96.4%) lasted less than 30 seconds. Only 1 alarm was acknowledged in the electronic health record. None of the alarms were clinically actionable or led to a code blue or death.

CONCLUSIONS

True AVR alarms may contribute to alarm fatigue. Hospitals should reevaluate the need for close monitoring of AVR and consider configuring this alarm to an inaudible message setting to reduce the risk of patient harm due to alarm fatigue. Prospective studies involving larger patient samples and varied monitors are warranted.

Для цитирования: Suba Sukardi, Sandoval Cass Piper, Zègre-Hemsey Jessica K, Hu Xiao, Pelter Michele M. Contribution of Electrocardiographic Accelerated Ventricular Rhythm Alarms to Alarm Fatigue.. American journal of critical care : an official publication, American Association of Critical-Care Nurses. 2019. DOI: 10.4037/ajcc2019314.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.