DOI 10.1227/neu.0000000000000053

Медицинская библиотека · Neurosurgery · 2013

Neuroform intracranial stenting for aneurysms using simple and multi-stent technique is associated with low risk of magnetic resonance diffusion-weighted imaging lesions.

Heller Robert S, Dandamudi Venkata, Calnan Daniel, Malek Adel M

СтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

BACKGROUND

Detection of procedural thromboembolism by diffusion-weighted magnetic resonance imaging (MR-DWI+) can help identify and mitigate endovascular risk factors. Data remain scant on procedural MR-DWI+ following the use of the Neuroform open-cell design stent in aneurysm embolization.

OBJECTIVE

We sought to evaluate the incidence of MR-DWI+ in Neuroform simple and multi-stent construct stenting for intracranial aneurysms in an attempt to delineate baseline risk and identify possible associated procedural factors.

METHODS

Seventy-six patients receiving 97 Neuroform stents in the treatment of intracranial aneurysm were identified from a prospective database and eligible for inclusion in the study. Diffusion-weighted magnetic resonance imaging (MR-DWI) was obtained in all patients within 48 hours of the procedure and reviewed for the presence of MR-DWI+ with patient records reviewed for analysis of factors predisposing to these lesions.

RESULTS

Patients were treated with single-stent Neuroform constructs in 57 cases (73%) and multi-stent Neuroform constructs in 21 cases (27%). Y-stent technique was used in 16 cases. MR-DWI+ was identified in 7 of 78 cases (9.0%), with MR-DWI+ in 0 of 10 subarachnoid hemorrhage cases. No MR-DWI lesions led to a permanent neurological deficit at discharge. There was no MR-DWI+ in patients treated with Y-stenting or multi-stent Neuroform constructs. The only factor associated with ipsilateral MR-DWI+ was target aneurysm location on an arterial sidewall over bifurcation (P = .01).

CONCLUSION

The Neuroform stent carries a very low risk of MR-DWI+ compared with its closed-cell design counterpart. Subarachnoid hemorrhage and deployment of multiple stents in the same anatomical region in configurations such as the Y-stent construct did not increase the risk of acute procedural thromboembolism.

Для цитирования: Heller Robert S, Dandamudi Venkata, Calnan Daniel, Malek Adel M. Neuroform intracranial stenting for aneurysms using simple and multi-stent technique is associated with low risk of magnetic resonance diffusion-weighted imaging lesions.. Neurosurgery. 2013. DOI: 10.1227/neu.0000000000000053.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.