DOI 10.1093/jalm/jfag031

Медицинская библиотека · The journal of applied laboratory medicine · 2026

Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 Assay.

Knoll Marko, Daniels Lori B, Mueller Christian, Mills Nicholas L, Giannitsis Evangelos, Rösser Alisa F A, Kurtoic Dunja, Wahl Annika, Body Richard, Christenson Robert H, Cobbaert Christa, deFilippi Christopher R, Eggers Kai M, Inoue Kenji, Jaffe Allan S, McCarthy Cian P, McCord James, Neumann Johannes T, Omland Torbjørn, Papendick Cynthia, Sandoval Yader, Tan Jack Wei Chieh, Than Martin P, Twerenbold Raphael, Peacock W Frank, Meex Steven J R

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

High-sensitivity cardiac troponin (hs-cTn) assays are recommended for the diagnosis of acute myocardial infarction. Here, we characterize the analytical performance of a next-generation hs-cTn assay, Elecsys® Troponin T hs Gen 6 (Roche Diagnostics International).

METHODS

Surplus lithium-heparin plasma or serum samples from patients or healthy volunteers were run on Cobas® e 801, e 402, and Pro analyzers. Limits of blank (LoB), limits of detection (LoD), and limits of quantitation (LoQ) were determined according to CLSI EP17-A2, with target values of 1.0 and 1.5 ng/L for LoB/LoD and 3.0 ng/L (10% CV) and 1.5 ng/L (20% CV) for LoQ, respectively. Precision was measured, per CLSI EP17-A2, using 3 QC samples (approximately 4, 30, and 220 ng/L), 12 native samples, and 3 reagent lots. Linearity, per CLSI EP06-Ed2, was determined by diluting samples with cardiac troponin T (cTnT) concentration above the measuring range with a low/blank sample. Interference (per Glick) with endogenous and assay components at 5 cTnT concentrations was assessed.

RESULTS

Measured values for LoB, LoD, and LoQ at 10% and 20% CV were 0.1 to 0.7 ng/L, 0.3 to 1.4 ng/L, 1.0 to 2.9 ng/L, and 0.4 to 1.2 ng/L, respectively. Repeatability CVs were 1.0 to 5.8% for mean cTnT concentrations of 2.6 to 9230 ng/L in lithium-heparin plasma. High precision was shown across lots, and linearity was observed across the measuring range (1.5 to 9500 ng/L, all Pearson's r = 1.00). No interferences were observed, specified up to ≤1000 mg/dL hemoglobin, ≤50 mg/dL [≤855 µmol/L] icterus/bilirubin, and ≤1200 ng/mL biotin.

CONCLUSIONS

The analytical performance characterization of the assay demonstrated high sensitivity, high precision at the low end and across the measuring range, and resistance to interference.

Для цитирования: Knoll Marko, Daniels Lori B, Mueller Christian, Mills Nicholas L, Giannitsis Evangelos, Rösser Alisa F A, Kurtoic Dunja, Wahl Annika, Body Richard, Christenson Robert H, Cobbaert Christa, deFilippi Christopher R, Eggers Kai M, Inoue Kenji, Jaffe Allan S, McCarthy Cian P, McCord James, Neumann Johannes T, Omland Torbjørn, Papendick Cynthia, Sandoval Yader, Tan Jack Wei Chieh, Than Martin P, Twerenbold Raphael, Peacock W Frank, Meex Steven J R. Analytical Performance Evaluation of the Cardiac Troponin T High Sensitivity Gen 6 Assay.. The journal of applied laboratory medicine. 2026. DOI: 10.1093/jalm/jfag031.