Bioprosthetic Total Artificial Heart in Autoregulated Mode Is Biologically Hemocompatible: Insights for Multimers of von Willebrand Factor.

Poitier Bastien, Chocron Richard, Peronino Christophe, Philippe Aurélien, Pya Yuri, Rivet Nadia, Richez Ulysse, Bekbossynova Mahabbat, Gendron Nicolas, Grimmé Marc, Bories Marie Cécile, Brichet Julie, Capel Antoine, Rancic Jeanne, Vedie Benoit, Roussel Jean Christian, Jannot Anne-Sophie, Jansen Piet, Carpentier Alain, Ivak Peter, Latremouille Christian, Netuka Ivan, Smadja David M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Carmat bioprosthetic total artificial heart (Aeson; A-TAH) is a pulsatile and autoregulated device. The aim of this study is to evaluate level of hemolysis potential acquired von Willebrand syndrome after A-TAH implantation.

METHODS

We examined the presence of hemolysis and acquired von Willebrand syndrome in adult patients receiving A-TAH support (n=10) during their whole clinical follow-up in comparison with control subjects and adult patients receiving Heartmate II or Heartmate III support. We also performed a fluid structure interaction model coupled with computational fluid dynamics simulation to evaluate the A-TAH resulting shear stress and its distribution in the blood volume.

RESULTS

The cumulative duration of A-TAH support was 2087 days. A-TAH implantation did not affect plasma free hemoglobin over time, and there was no association between plasma free hemoglobin and cardiac output or beat rate. For VWF (von Willebrand factor) evaluation, A-TAH implantation did not modify multimers profile of VWF in contrast to Heartmate II and Heartmate III. Furthermore, fluid structure interaction coupled with computational fluid dynamics showed a gradually increase of blood damage according to increase of cardiac output (P<0.01), however, the blood volume fraction that endured significant shear stresses was always inferior to 0.03% of the volume for both ventricles in all regimens tested. An inverse association between cardiac output, beat rate, and high-molecular weight multimers ratio was found.

CONCLUSIONS

We demonstrated that A-TAH does not cause hemolysis or AWVS. However, relationship between HMWM and cardiac output depending flow confirms relevance of VWF as a biological sensor of blood flow, even in normal range.

Для цитирования: Poitier Bastien, Chocron Richard, Peronino Christophe, Philippe Aurélien, Pya Yuri, Rivet Nadia, Richez Ulysse, Bekbossynova Mahabbat, Gendron Nicolas, Grimmé Marc, Bories Marie Cécile, Brichet Julie, Capel Antoine, Rancic Jeanne, Vedie Benoit, Roussel Jean Christian, Jannot Anne-Sophie, Jansen Piet, Carpentier Alain, Ivak Peter, Latremouille Christian, Netuka Ivan, Smadja David M. Bioprosthetic Total Artificial Heart in Autoregulated Mode Is Biologically Hemocompatible: Insights for Multimers of von Willebrand Factor.. Arteriosclerosis, thrombosis, and vascular biology. 2022. DOI: 10.1161/atvbaha.121.316833.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.