[Mejora progresiva de los umbrales de captura auricular y ventricular, detección e impedancias en los cables de estimulación epicárdica en adultos jóvenes que se someten a conversión de Fontan].

Moltedo Jose M, Abello Mauricio S, Doiny David, Falconi Estela C, Diaz Carlos J, Majdalani María G, Cannon Bryan C

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Patients with univentricular hearts who require permanent pacing systems typically require placement of epicardial leads. It is frequently difficult to find a position with good thresholds due to epimyocardial fibrosis or fat. The goal of the study is to assess the progression of capture thresholds (CT), sensing parameters (P waves and R waves), and impedances (imp) of steroid eluting epicardial pacing leads in young adults who underwent Fontan conversion and a pacemaker implant.

METHODS

All patients undergoing Fontan conversion in two institutions were retrospectively identified. Demographic data, congenital heart defects, pacing leads used, and pacing parameters were analyzed at implant, at 6 weeks and 12 months after implant.

RESULTS

Twenty patients were identified (twelve males); mean age at conversion was 24.9 ± 5.4 years (range 18-35). Epicardial bipolar steroid eluting leads were used. The site of implant both in the atria and the ventricles varied depending on the parameters. At implant, mean atrial and ventricular impedances were 617 ± 171 Ω and 1061 ± 771 Ω, respectively, mean P wave amplitude was 2 ± 0.7 mV, and mean R wave amplitude was 12.5 ± 7.7 mV. Mean CT was 1.7 ± 0.8 V at 0.5 ms for the atrium and 2.2 ± 1.2 V at 0.5 ms for the ventricle. Ventricular CT and impedance showed an improvement within the first 12 months after implant, with four patients having a decrease in threshold of more than 2 V.

CONCLUSION

In patients undergoing Fontan conversion, implant ventricular CT and impedances are frequently higher than expected but typically improve during follow-up. Acceptance of higher initial threshold values may be a potential strategy in this patient population.

Для цитирования: Moltedo Jose M, Abello Mauricio S, Doiny David, Falconi Estela C, Diaz Carlos J, Majdalani María G, Cannon Bryan C. [Mejora progresiva de los umbrales de captura auricular y ventricular, detección e impedancias en los cables de estimulación epicárdica en adultos jóvenes que se someten a conversión de Fontan].. Archivos de cardiologia de Mexico. 2021. DOI: 10.24875/acm.20000426.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.