DOI 10.1097/j.jcrs.0000000000001443

Медицинская библиотека · Journal of cataract and refractive surgery · 2024

Repeatability of pyramidal aberrometer measurements in keratoconus and normal eyes.

Ibrahim Perla, Assaf Jad F, Bejjani Romy, Torbey Julien, Yehia Madeleine, Bahir Al-Ulloom Saleh, Awwad Shady T

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the intrasubject repeatability of pyramidal aberrometer measurements in a sample of keratoconus and normal eyes.

SETTING

American University of Beirut Medical Center, Beirut, Lebanon.

DESIGN

Prospective comparative repeatability analysis.

METHODS

Study population: Keratoconus and normal eyes from adult patients. Observation procedures: Each eye was evaluated with 3 consecutive acquisitions using a pyramidal aberrometer. Main outcome measures: The repeatability of different ocular higher-order aberrations and lower-order aberrations (HOAs and LOAs, respectively), and Zernike coefficients down to the fifth order, was evaluated. Repeatability was assessed by within-subject SDs (Sw), repeatability limits ( r ), and intraclass correlation coefficients (ICCs), among other parameters.

RESULTS

72 keratoconus patients (72 eyes) and 76 normal patients (76 eyes) were included. In normal and keratoconus eyes, the ICC of total LOAs and HOAs, as well as each of the Zernike coefficients, was >0.9. The Sw for keratoconus eyes with mean maximal keratometry (Kmax) <50 diopters (D) was 0.1345 for total LOAs, 0.0619 for total HOAs, 0.0292 for horizontal coma, 0.0561 for vertical coma, and 0.0221 for spherical aberration as compared with 0.2696, 0.1486, 0.0972, 0.1497, and 0.0757 for keratoconus eyes with Kmax ≥50 D. Similar trend of better repeatability for grade 1 keratoconus and HOAs <2 D as compared with grades 2 and 3 keratoconus and eyes with HOAs >2 D were also noted.

CONCLUSIONS

Ocular aberrometer measurements generated by high definition pyramidal aberrometers have high repeatability in both normal and mild keratoconus eyes and moderate repeatability, yet still clinically acceptable, in advanced keratoconus. This is of particular importance in ocular wavefront-guided treatments.

Для цитирования: Ibrahim Perla, Assaf Jad F, Bejjani Romy, Torbey Julien, Yehia Madeleine, Bahir Al-Ulloom Saleh, Awwad Shady T. Repeatability of pyramidal aberrometer measurements in keratoconus and normal eyes.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001443.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.