Subclinical Keratoconus Detection and Characterization Using Motion-Tracking Brillouin Microscopy.

Randleman J Bradley, Zhang Hongyuan, Asroui Lara, Tarib Imane, Dupps William J, Scarcelli Giuliano

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To characterize focal biomechanical alterations in subclinical keratoconus (SKC) using motion-tracking (MT) Brillouin microscopy and evaluate the ability of MT Brillouin metrics to differentiate eyes with SKC from normal control eyes.

DESIGN

Prospective cross-sectional study.

PARTICIPANTS

Thirty eyes from 30 patients were evaluated, including 15 eyes from 15 bilaterally normal patients and 15 eyes with SKC from 15 patients.

METHODS

All patients underwent Scheimpflug tomography and MT Brillouin microscopy using a custom-built device. Mean and minimum MT Brillouin values within the anterior plateau region and anterior 150 μm were generated. Scheimpflug metrics evaluated included inferior-superior (IS) value, maximum keratometry (Kmax), thinnest corneal thickness, asymmetry indices, Belin/Ambrosio display total deviation, and Ambrosio relational thickness. Receiver operating characteristic (ROC) curves were generated for all Scheimpflug and MT Brillouin metrics evaluated to determine the area under the ROC curve (AUC), sensitivity, and specificity for each variable.

MAIN OUTCOME MEASURES: Discriminative performance based on AUC, sensitivity, and specificity.

RESULTS

No significant differences were found between groups for age, sex, manifest refraction spherical equivalent, corrected distance visual acuity, Kmax, or KISA% index. Among Scheimpflug metrics, significant differences were found between groups for thinnest corneal thickness (556 μm vs. 522 μm; P < 0.001), IS value (0.29 diopter [D] vs. 1.05 D; P < 0.001), index of vertical asymmetry (IVA; 0.10 vs. 0.19; P < 0.001), and keratoconus index (1.01 vs. 1.05; P < 0.001), and no significant differences were found for any other Scheimpflug metric. Among MT Brillouin metrics, clear differences were found between control eyes and eyes with SKC for mean plateau (5.71 GHz vs. 5.68 GHz; P < 0.0001), minimum plateau (5.69 GHz vs. 5.65 GHz; P < 0.0001), mean anterior 150 μm (5.72 GHz vs. 5.68 GHz; P < 0.0001), and minimum anterior 150 μm (5.70 GHz vs. 5.66 GHz; P < 0.001). All MT Brillouin plateau and anterior 150 μm mean and minimum metrics fully differentiated groups (AUC, 1.0 for each), whereas the best performing Scheimpflug metrics were keratoconus index (AUC, 0.91), IS value (AUC, 0.89), and IVA (AUC, 0.88).

CONCLUSIONS

Motion-tracking Brillouin microscopy metrics effectively characterize focal corneal biomechanical alterations in eyes with SKC and clearly differentiated these eyes from control eyes, including eyes that were not differentiated accurately using Scheimpflug metrics.

FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Для цитирования: Randleman J Bradley, Zhang Hongyuan, Asroui Lara, Tarib Imane, Dupps William J, Scarcelli Giuliano. Subclinical Keratoconus Detection and Characterization Using Motion-Tracking Brillouin Microscopy.. Ophthalmology. 2023. DOI: 10.1016/j.ophtha.2023.10.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.