Epithelial Thickness Mapping in Keratoconic Corneas: Repeatability and Agreement Between CSO MS-39, Heidelberg Anterion, and Optovue Avanti OCT Devices.

Feng Yue, Reinstein Dan Z, Nitter Tore, Archer Timothy J, McAlinden Colm, Bertelsen Geir, Stojanovic Aleksandar

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess repeatability and agreement of corneal epithelial thickness mapping in eyes with keratoconus using three optical coherence tomography (OCT) devices featuring different technologies: spectral-domain (SD) OCT combined with Placido disk corneal topography (MS-39), swept-source OCT (Anterion), and SD-OCT (Avanti).

METHODS

Three consecutive measurements were acquired with the three devices in 60 eyes with keratoconus. The mean epithelial thickness was calculated in the central 2-mm zone and in 2- to 5-mm and 5- to 7-mm diameter rings. The repeatability was calculated using pooled within-subject standard deviation (Sw). The agreement was assessed by paired t tests and Bland-Altman plots.

RESULTS

The repeatability (Sw) of the epithelial thickness for the central 2-mm zone was 0.91, 0.71, and 0.93 μm for the MS-39, Anterion, and Avanti, respectively. All thicknesses with the MS-39 were greater than those of the Anterion and Avanti, with mean differences of 4.11 ± 1.34 μm (P < .001) and 0.52 ± 1.30 μm (P = .003), respectively. The 95% limits of agreement were 1.484 to 6.736 μm for the MS-39 and Anterion, -3.068 to 2.028 μm for the Avanti and MS-39, and 1.258 to 5.922 μm for for the Avanti and Anterion.

CONCLUSIONS

Epithelial thickness mapping results were most repeatable with the Anterion, followed by the MS-39 and Avanti. The MS-39 gave the thickest values, followed by the Avanti and Anterion. The differences were significant, making the devices not interchangeable for epithelial thickness mapping in eyes with keratoconus. [J Refract Surg. 2023;39(7):474-480.].

Для цитирования: Feng Yue, Reinstein Dan Z, Nitter Tore, Archer Timothy J, McAlinden Colm, Bertelsen Geir, Stojanovic Aleksandar. Epithelial Thickness Mapping in Keratoconic Corneas: Repeatability and Agreement Between CSO MS-39, Heidelberg Anterion, and Optovue Avanti OCT Devices.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2023. DOI: 10.3928/1081597x-20230606-01.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.