DOI 10.1097/ftd.0000000000001236
Development and Validation of a Sensitive Liquid Chromatography-Tandem Mass Spectrometry Method for Therapeutic Drug Monitoring of Ceftolozane and Tazobactam in Human Plasma Microsamples.
Аннотация
BACKGROUND
Ceftolozane/tazobactam (C-T) is a novel beta-lactam/beta-lactamase inhibitor combination approved for the treatment of various infections caused by difficult-to-treat Pseudomonas aeruginosa . In critically ill patients, C-T may exhibit significant pharmacokinetic variability, both between individuals and within individuals, warranting therapeutic drug monitoring for clinical purposes. We aim to develop and validate a novel and sensitive analytical method for concurrently determining C and T in human plasma microsamples (3 μL).
METHODS
The method was developed using liquid chromatography-tandem mass spectrometry (LC-MS/MS) with positive electrospray ionization and multiple reaction monitoring (MRM) detection modes, employing specific mass transitions for both drugs. Sample preparation was simple, and the chromatographic run lasted only 4 minutes. Validation was conducted according to European Medicines Agency (EMA) guidelines, encompassing specificity, sensitivity, linearity, precision, accuracy, matrix effect, extraction recovery, limit of quantification, and drug stability. The validated method was applied to measure C and T in 32 plasma samples collected from critically ill patients with multidrug-resistant, gram-negative, bacterial infections.
RESULTS
The method ensured accurate (BIAS% 2.1-9.6 for C and -2.2 to 15.2 for T) and precise intraday CV% for C: 6.7-5.5; for T: 1.3-8.9; interday CV% for C 6.0-10.8; for T 4.1-10.2) measurements of C-T over a wide concentration range (0.2-200.0 mg/L for C and 0.1-100.0 mg/L for T). Overall, the recovery at quality control concentration levels was high for both C and T (mean values: 90-91 for C and 89-92 for T). Analyte stability was satisfactory in both human plasma and extracts under various storage conditions. The clinical applicability of the assay was confirmed by the reliably quantifying C and T in clinical plasma samples.
CONCLUSIONS
The developed and validated LC-MS/MS method is sensitive and suitable for monitoring C and T in human plasma microsamples.