DOI 10.1016/j.ijcard.2023.131597
About different localization of hypoattenuated lesions following transcatheter aortic valve replacement.
Аннотация
BACKGROUND
Subclinical leaflet thrombosis is diagnosed using multidetector computed tomography (MDCT) and is characterised by a meniscal-shaped hypoattenuated lesion of one or more leaflets. Transcatheter aortic self-expandable valves are commonly manufactured with pliable pericardium over a nitinol frame that forms leaflet and extra-leaflet components such as the valve skirt. Little is known about extra-leaflet hypoattenuated lesion localisation, including that at the anatomical sinus level. Thus, the main aim of this study was to describe leaflet and extra-leaflet (anatomic sinus and subvalvular level) hypoattenuated lesions following transcatheter aortic valve replacement with a self-expandable prosthesis. As a secondary aim, we sought to investigate predictors of hypoattenuated lesions.
METHODS
Fifty patients underwent MDCT at the follow-up.
RESULTS
At a follow-up of 12 months, hypoattenuated leaflet lesions with mild to severe restricted movement were detected in eight individuals (16%), anatomic sinus lesions were identified in nine patients (18%), with higher prevalence in the non-coronary sinus (16%), and subvalvular lesions with variable extension toward the valve inflow were diagnosed in eight patients (16%). In 4 patients (8%) the anatomic sinus thrombus was 'in overlap' with leaflet thrombus; in 3 patients (6%) was in continuity with subvalvular frame thrombus. Bicuspid valve was the only independent predictor associated with hypoattenuated lesions (adj OR 8.25 (95% CI: 1.38, 49.21), p = 0.02)).
CONCLUSIONS
This study demonstrated that hypoattenuated lesions could be identified not only at the leaflet but also at the subvalvular and anatomic sinus levels. The clinical relevance of such lesions remains unclear.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.