Bachmann's Bundle Pacing not Only Improves Interatrial Conduction but Also Reduces the Need for Ventricular Pacing.
Аннотация
BACKGROUND
Patients treated for sick sinus syndrome may have interatrial conduction disorder leading to atrial fibrillation.
OBJECTIVES
This study was aimed to assess the influence of the atrial pacing site on interatrial and atrioventricular conduction as well as the percentage of ventricular pacing in patients with sick sinus syndrome implanted with atrioventricular pacemaker.
MATERIAL AND
METHODS
The study population: 96 patients (58 females, 38 males) aged 74.1 ± 11.8 years were divided in two groups: Group 1 (n = 44) with right atrial appendage pacing and group 2 (n = 52) with Bachmann's area pacing. We assessed the differences in atrioventricular conduction in sinus rhythm and atrial 60 and 90 bpm pacing, P-wave duration and percentage of ventricular pacing.
RESULTS
No differences in baseline P-wave duration in sinus rhythm between the groups (102.4 ± 17 ms vs. 104.1 ± 26 ms, p = ns.) were noted. Atrial pacing 60 bpm resulted in longer P-wave in group 1 vs. group 2 (138.3 ± 21 vs. 106.1 ± 15 ms, p < 0.01). The differences between atrioventricular conduction time during sinus rhythm and atrial pacing at 60 and 90 bpm were significantly longer in patients with right atrial appendage vs. Bachmann's pacing (44.1 ± 17 vs. 9.2 ± 7 ms p < 0.01 and 69.2 ± 31 vs. 21.4 ± 12 ms p < 0.05, respectively). The percentage of ventricular pacing was higher in group 1 (21 vs. 4%, p < 0.01).
CONCLUSIONS
Bachmann's bundle pacing decreases interatrial and atrioventricular conduction delay. Moreover, the frequency-dependent atrioventricular conduction lengthening is much less pronounced during Bachmann's bundle pacing. Right atrial appendage pacing in sick sinus syndrome patients promotes a higher percentage of ventricular pacing.
Предметные рубрики MeSH