Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study.

Ritter Philippe, Duray Gabor Z, Steinwender Clemens, Soejima Kyoko, Omar Razali, Mont Lluís, Boersma Lucas V A, Knops Reinoud E, Chinitz Larry, Zhang Shu, Narasimhan Calambur, Hummel John, Lloyd Michael, Simmers Timothy Alexander, Voigt Andrew, Laager Verla, Stromberg Kurt, Bonner Matthew D, Sheldon Todd J, Reynolds Dwight, Micra Transcatheter Pacing Study Group

Оценочное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: Permanent cardiac pacing is the only effective treatment for symptomatic bradycardia, but complications associated with conventional transvenous pacing systems are commonly related to the pacing lead and pocket. We describe the early performance of a novel self-contained miniaturized pacemaker.

METHODS AND

RESULTS

Patients having Class I or II indication for VVI pacing underwent implantation of a Micra transcatheter pacing system, from the femoral vein and fixated in the right ventricle using four protractible nitinol tines. Prespecified objectives were >85% freedom from unanticipated serious adverse device events (safety) and <2 V 3-month mean pacing capture threshold at 0.24 ms pulse width (efficacy). Patients were implanted (n = 140) from 23 centres in 11 countries (61% male, age 77.0 ± 10.2 years) for atrioventricular block (66%) or sinus node dysfunction (29%) indications. During mean follow-up of 1.9 ± 1.8 months, the safety endpoint was met with no unanticipated serious adverse device events. Thirty adverse events related to the system or procedure occurred, mostly due to transient dysrhythmias or femoral access complications. One pericardial effusion without tamponade occurred after 18 device deployments. In 60 patients followed to 3 months, mean pacing threshold was 0.51 ± 0.22 V, and no threshold was ≥2 V, meeting the efficacy endpoint (P < 0.001). Average R-wave was 16.1 ± 5.2 mV and impedance was 650.7 ± 130 ohms.

CONCLUSION

Early assessment shows the transcatheter pacemaker can safely and effectively be applied. Long-term safety and benefit of the pacemaker will further be evaluated in the trial.

CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov ID NCT02004873.

Для цитирования: Ritter Philippe, Duray Gabor Z, Steinwender Clemens, Soejima Kyoko, Omar Razali, Mont Lluís, Boersma Lucas V A, Knops Reinoud E, Chinitz Larry, Zhang Shu, Narasimhan Calambur, Hummel John, Lloyd Michael, Simmers Timothy Alexander, Voigt Andrew, Laager Verla, Stromberg Kurt, Bonner Matthew D, Sheldon Todd J, Reynolds Dwight, Micra Transcatheter Pacing Study Group. Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study.. European heart journal. 2015. DOI: 10.1093/eurheartj/ehv214.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.