MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.medici.2014.11.010

Медицинская библиотека · Medicina (Kaunas, Lithuania) · 2014

Evaluation of left ventricular longitudinal function and synchrony after dual chamber pacemaker implantation.

Baronaitė-Dūdonienė Kristina, Vaškelytė Jolanta, Puodžiukynas Aras, Zabiela Vytautas, Kazakevičius Tomas, Šakalytė Gintarė

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

To evaluate left ventricular (LV) longitudinal function and dyssynchrony mechanisms after dual chamber pacemaker implantation.

MATERIALS AND

METHODS

The speckle tracking imaging technique was used for quantification of global longitudinal function of the left ventricle and for dyssynchrony evaluation before pacemaker implantation and after 3-month follow-up. The study group consisted of 98 patients with conventional indications for dual chamber pacemaker implantation.

RESULTS

Speckle tracking echocardiographic methods and image postprocessing revealed impairment of global longitudinal strain and significant LV dyssynchrony derived from 12 basal and mid-septum segments usually untraceable with conventional echocardiographic methods. Despite good physical performance and ejection fraction, global longitudinal strain significantly decreased in all patients from -15.08±0.46 to -13.56±0.5 (P<0.05) as well as mitral annulus movement decreased from 11.57±2.41 to 8.46±1.74cm/s (P<0.001) and from 12.55±2.75 to 10.78±2.82mm (P<0.001). It was expected that patients with dual chamber pacemaker will develop inter- and intraventricular dyssynchrony, but our study showed that pacing lead position did not prevent from LV dysynchronisation and only changed the mechanism.

CONCLUSIONS

Global longitudinal strain and LV dyssynchrony assessment enables us to detect early signs of LV dysfunction. Mechanisms of dyssynchrony development will be useful for pacemaker programing choices in order to prevent further dyssynchronisation.

Для цитирования: Baronaitė-Dūdonienė Kristina, Vaškelytė Jolanta, Puodžiukynas Aras, Zabiela Vytautas, Kazakevičius Tomas, Šakalytė Gintarė. Evaluation of left ventricular longitudinal function and synchrony after dual chamber pacemaker implantation.. Medicina (Kaunas, Lithuania). 2014. DOI: 10.1016/j.medici.2014.11.010.