P-wave evidence as a method for improving algorithm to detect atrial fibrillation in insertable cardiac monitors.

Pürerfellner Helmut, Pokushalov Evgeny, Sarkar Shantanu, Koehler Jodi, Zhou Ren, Urban Lubos, Hindricks Gerhard

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Frequent premature atrial contractions and sick sinus syndrome are primary causes of inappropriate atrial fibrillation (AF) detection in insertable cardiac monitors (ICMs).

OBJECTIVE

The study aimed to validate an algorithm designed to reduce inappropriate AF detection on the basis of the identification of a single P wave during the cardiac cycle.

METHODS

The original detection algorithm looks for evidence of AF based on differences in the pattern of R-R intervals over a 2-minute period. The improved algorithm reduces evidence for AF detection if P waves are detected. The algorithm was validated by using Holter data, which collected 2 leads of surface electrocardiogram and continuously uplinked ICM electrocardiogram over a 46-hour period. ICM detections were compared with Holter annotations to compute episode and duration detection performance.

RESULTS

Valid Holter recordings (8442 hours) were analyzed from 206 patients. True AF was observed in 76 patients, yielding 482 true AF episodes ≥2 minutes in duration and 1191 hours of AF. The algorithm correctly identified 97.8% of the total AF duration and 99.3% of the total sinus or non-AF rhythm duration. The algorithm detected 85% (90% per-patient average) of all AF episodes ≥2 minutes in duration, and 55% (78% per-patient average) of the detected episodes had AF. AF was found in 95% of the detected episodes >1 hour. The improved algorithm reduced inappropriate episodes and duration by 46% and 55%, respectively, while also reducing appropriate episodes and duration by 2% and 0.1%, respectively.

CONCLUSION

An improvement in the ICM algorithm for AF detection incorporating P-wave information substantially reduced inappropriately detected episodes and duration, with minimal reduction in sensitivity for detecting AF.

Для цитирования: Pürerfellner Helmut, Pokushalov Evgeny, Sarkar Shantanu, Koehler Jodi, Zhou Ren, Urban Lubos, Hindricks Gerhard. P-wave evidence as a method for improving algorithm to detect atrial fibrillation in insertable cardiac monitors.. Heart rhythm. 2014. DOI: 10.1016/j.hrthm.2014.06.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.