Potential prevention of pacing-induced heart failure using simple pacemaker programming algorithm.
Аннотация
INTRODUCTION
Right ventricular pacing (RVP) causes ventricular desynchronization and may lead to the development of heart failure (HF). Prolongation of atrioventricular delay (AVD) in DDDR pacemakers reduces unnecessary RV stimulation. The aim of the study was to verify the influence of RVP reduction on HF symptoms.
METHODS
The study comprised 31 patients (17 men, mean age: 71.6 ± 8 yrs) with DDDR pacemaker implanted due to sinus node dysfunction (SND). At baseline, 28 patients did not present any symptoms of HF. Three patients were in NYHA class II. Patients were randomized either to 150 ms AVD or to minimizing right ventricular pacing (MRVP). Crossing over to the alternate mode took place after 4 months. Cardiopulmonary exercise test (CPX), echocardiography (ECHO) and BNP measurements were done before pacemaker implantation, after 4 and 8 months.
RESULTS
The percentage of RVP was significantly higher in 150 ms AVD than in MRVP: 81.7 ± 22.6 versus 14.2±20.5%, P < 0.0001. Patients with 150 ms mode had worse CPX parameters than those with MRVP mode: peak oxygen uptake was 14.2±4.3 versus 19.9±6.3 ml/kg per min, P = 0.0001, higher BNP concentrations: 72.3±48.3 versus 49.4±43.9 pg/ml, P = 0.001 and worse left ventricle [LV] function: ejection fraction: 53.2±6.7 versus 57.3±5.5%, P < 0.0001; LV diastolic diameter: 4.86±0.52 versus 4.66±0.5 cm, P < 0.01.
CONCLUSION
Predominant RVP in patients without symptoms of HF at baseline may be responsible for worse performance in cardiopulmonary exercise test, higher BNP concentrations and impairment of LV function. Specific DDDR pacemaker programming promotes intrinsic AV conduction and may prevent the development of pacing-induced HF.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.