What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?

Arnholt Christina M, MacDonald Daniel W, Klein Gregg R, Cates Harold E, Rimnac Clare M, Kurtz Steven M, Implant Research Center Writing Committee, Kocagoz Sevi, Chen Antonia F

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The purpose of this study was to determine the incidence of metal release in contemporary total knee arthroplasty and the patient-related factors associated with this release.

METHODS

In total, 256 retrieved cobalt-chromium femoral components were collected through a multi-institutional orthopedic implant retrieval program (implanted: 1-15 years). Implants were mainly revised for loosening (84/256), instability (62/256), and infection (46/256). Third-body damage was assessed using a semiquantitative scoring method. Microscale electro-corrosion damage (MECD) was evaluated using digital optical microscopy. Radii of curvature were measured from representative components to calculate anterior-posterior and medial-lateral ratios. Femoral component surface roughness was measured using a white light interferometer. Using a multivariable linear model, associations between damage score, implant, and patient factors were tested. Spearman's ρ correlation tests were performed to determine the association between roughness measurements and damage score.

RESULTS

Mild to severe damage was observed in 52% (134/256) of the components. In the multivariable linear model, anterior-posterior ratio (β = -8.07; P < .001), loosening (β = -0.52; P = .006), and patient weight (β = 0.01; P = .007) were associated with damage score. Suspected MECD damage was observed in 82% (209/256) of components. The Ra value (ρ = 0.196; P = .002) and Rq value (ρ = 0.157; P = .012) increased as the damage score increased.

CONCLUSION

The findings of this retrieval study support that similar damage mechanisms exist in contemporary and long-term total knee arthroplasty devices. Additionally, we observed associations between loosening, anterior-posterior conformity, and patient weight with increased surface damage.

Для цитирования: Arnholt Christina M, MacDonald Daniel W, Klein Gregg R, Cates Harold E, Rimnac Clare M, Kurtz Steven M, Implant Research Center Writing Committee, Kocagoz Sevi, Chen Antonia F. What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?. The Journal of arthroplasty. 2018. DOI: 10.1016/j.arth.2018.05.020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.