A multicenter clinical evaluation of the Clot Signature Analyzer.
Аннотация
BACKGROUND
The Clot Signature Analyzer (CSA) was designed to assess global hemostasis as a screening assay using non-anticoagulated whole blood. Three different measurements are produced by the instrument: platelet hemostasis time (PHT), clot time (CT), and collagen-induced thrombus formation (CITF).
OBJECTIVES
The purpose of the present study was to determine normal ranges for these measurements and assess the performance of the CSA in patients with well-characterized hemostatic disorders and in normal subjects.
PATIENTS AND
METHODS
Four institutions participated in the study. Each established their own normal reference ranges. Patients with well-characterized hemostatic disorders and concurrent normal controls were subsequently examined.
RESULTS
Normal ranges between institutions were similar although statistically different. One hundred and eight patients were examined: 46 individuals with von Willebrand disease (VWD) (type 1, 26; type 2A, 11; type 2B, six; type 3, three); 38 patients with a coagulation factor deficiency; 13 individuals with platelet function defects; 10 patients taking warfarin; and one individual on low-molecular-weight heparin. Of these patients, 89% had at least one abnormality by CSA: 42/46 VWD patients, 35/38 coagulation protein defect patients, 9/13 patients with platelet function defects, 9/10 patients on warfarin and 1/1 patient on low-molecular-weight heparin. Of 116 normal subjects, 103 (89%) fell within the centers' normal range. These data suggest that the CSA has a good sensitivity for bleeding disorders.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.