DOI 10.1016/j.acra.2026.06.054
CCTA without ECG Leads Using Whole-Heart Motion Correction Algorithm Based on Wide-Detector CT.
Аннотация
PURPOSE: This study evaluates the impact of a whole heart motion correction algorithm on image quality and interpretability in coronary computed tomography angiography (CCTA) without ECG leads.
MATERIALS AND
METHODS
Seventy patients with acute chest pain underwent CCTA without ECG leads examination. Images were reconstructed with standard (STD), coronary vessel tracking motion correction (Snapshot Freeze 1, SSF1), and whole‑heart motion correction (Snapshot Freeze 2, SSF2) algorithms. Additionally, all three groups underwent deep learning image reconstruction (DLIR) algorithm. Objective evaluation, including noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured. Subjective image quality was scored by two radiologists using a 5 - point Likert scale. Interpretability of coronary vessels was assessed per segment, vessel, and patient.
RESULTS
SSF2 significantly reduced noise and improved SNR/CNR compared to STD and SSF1 (all P < .05). Subjective scores were highest for SSF2 (median 5.0 vs 4.0 for SSF1 and 3.0 for STD, P < .001), with inter-observer agreement (κ = 0.87). SSF2 achieved higher interpretability for proximal/mid-segments (99.6%) and improved distal/small-branch interpretability (97.2% vs 89.4% for SSF1, P < .017). Patient-level interpretability reached 100% with SSF2, vs 92.9% (SSF1) and 51.4% (STD).
CONCLUSION
SSF2 enhances image quality and interpretability in CCTA without ECG leads, and exhibits favorable in the visualization of distal vessels and small branches.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.