Assessment of weight-based variations in effective dose for common PET/CT scans in a tertiary hospital.

Al Said Abdallah T, Rahman Azhar A, Albadarneh Laith K, Oglat Ammar A, Binti A Halim Fatin N, Firas Wesal, Alshloul Yahia, Radaideh Khaldoon M

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Positron Emission Tomography/Computed Tomography (PET/CT) is a crucial imaging modality widely used for diagnosing and staging various cancers. Although patient weight is known to influence radiation dose, there is limited research on optimizing PET/CT protocols based on weight categories to minimize unnecessary exposure.

OBJECTIVES

This study evaluates the effective dose for common PET/CT scans at Penang Hospital, with a particular focus on the influence of patient weight on both PET and CT dose components. Additionally, the study assesses whether weight-based dose protocols can contribute to dose optimization.

METHODS

This retrospective study analyzed 1439 PET/CT scans performed at Penang Hospital. The effective dose was calculated using International Commission on Radiological Protection (ICRP) 103, 106, and 128 dose coefficients. Conversion factors used were 0.019 mSv/MBq for 18F-FDG, 0.0199 mSv/MBq for 68Ga-PSMA, and 0.021 mSv/MBq for 68Ga-DOTATATE. The CT dose component was estimated using the dose-length product (DLP) with an empirical conversion factor (k = 0.015 mSv·mGy-1·cm-1) for whole-body CT.

RESULTS

The total effective dose for 18F-FDG (head-to-thigh and head-to-feet) scans was 12.34 and 16.38 mSv, respectively. For 68Ga-PSMA and 68Ga-DOTATATE scans, the total effective dose was 11.55 and 12.00 mSv, respectively. A statistically significant increase in dose was observed in heavier patients due to increased administered activity and CT parameters (P < .05). Weight had a smaller impact on PET dose in 68Ga-PSMA and 68Ga-DOTATATE compared to 18F-FDG.

CONCLUSION

Patient weight significantly influences the total effective dose in PET/CT examinations. These findings highlight the importance of considering patient-specific factors when evaluating radiation dose in clinical practice.

Для цитирования: Al Said Abdallah T, Rahman Azhar A, Albadarneh Laith K, Oglat Ammar A, Binti A Halim Fatin N, Firas Wesal, Alshloul Yahia, Radaideh Khaldoon M. Assessment of weight-based variations in effective dose for common PET/CT scans in a tertiary hospital.. Radiation protection dosimetry. 2026. DOI: 10.1093/rpd/ncag078.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.