Safety and efficacy of applying a low-dose radiation fluoroscopy protocol in device implantations.
Аннотация
AIMS: For cardiac implantable electronic device (CIED) implantations, visualization of lead placement is necessary and fluoroscopy remains by far the most commonly used technique. With simple changes in the X-ray system settings, total radiation dose can be reduced significantly. The purpose of this study was to assess the safety and efficacy of various CIED implantations performed after implementation of a new dose reduction protocol (DRP).
METHODS AND
RESULTS
We conducted a retrospective chart review of 584 patients undergoing CIED implantation or revision in our hospital. Of these patients, 280 (48%) underwent the implantation prior to and 304 (52%) after the DRP introduction. The DRP included various changes for optimized image processing and exposure system settings to enable dose reduction, as well as a reduced frame rates (4 FPS for fluoroscopy and 7.5 FPS for cinematographic images). Of the 584 patients, 53 (9.1%) had a one-chamber pacemaker, 232 (39.7%) a two-chamber pacemaker, 133 (22.8%) a one-chamber ICD, 35 (6.0%) a two-chamber ICD, 82 (14.0%) a CRT (de novo) implantation, and 49 (8.3%) had an upgrade to a CRT device. DRP was associated with a 64% reduction of the dose-area product (1372 ± 2659 vs. 3792 ± 5025 cGcm2, P < 0.001), while fluoroscopy duration (13 ± 15 vs. 13 ± 15 min) and procedural duration (93 ± 52 vs. 92 ± 52 min.) did not significantly increase. Complication rates did not differ significantly between the two groups.
CONCLUSION
The DRP proved to effectively reduce radiation dose for all types of CIED implantations. Fluoroscopy time, total procedure time, and the number of complications did not increase after introducing the DRP.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.