DOI 10.1136/bjophthalmol-2012-301525

Медицинская библиотека · The British journal of ophthalmology · 2013

The Argus II epiretinal prosthesis system allows letter and word reading and long-term function in patients with profound vision loss.

da Cruz Lyndon, Coley Brian F, Dorn Jessy, Merlini Francesco, Filley Eugene, Christopher Punita, Chen Fred K, Wuyyuru Varalakshmi, Sahel Jose, Stanga Paulo, Humayun Mark, Greenberg Robert J, Dagnelie Gislin, Argus II Study Group

Контролируемое клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Retinal prosthesis systems (RPS) are a novel treatment for profound vision loss in outer retinal dystrophies. Ideal prostheses would offer stable, long-term retinal stimulation and reproducible spatial resolution in a portable form appropriate for daily life.

METHODS

We report a prospective, internally controlled, multicentre trial of the Argus II system. Twenty-eight subjects with light perception vision received a retinal implant. Controlled, closed-group, forced-choice letter identification, and, open-choice two-, three- and four-letter word identification tests were carried out.

RESULTS

The mean±SD percentage correct letter identification for 21 subjects tested were: letters L, T, E, J, F, H, I, U, 72.3±24.6% system on and 17.7±12.9% system off; letters A, Z, Q, V, N, W, O, C, D, M, 55.0±27.4% system on and 11.8%±10.7% system off, and letters K, R, G, X, B, Y, S, P, 51.7±28.9% system on and 15.3±7.4% system off. (p<0.001 for all groups). A subgroup of six subjects was able to consistently read letters of reduced size, the smallest measuring 0.9 cm (1.7°) at 30 cm, and four subjects correctly identify unrehearsed two-, three- and four-letter words. Average implant duration was 19.9 months.

CONCLUSIONS

Multiple blind subjects fitted with the Argus II system consistently identified letters and words using the device, indicating reproducible spatial resolution. This, in combination with stable, long-term function, represents significant progress in the evolution of artificial sight.

Для цитирования: da Cruz Lyndon, Coley Brian F, Dorn Jessy, Merlini Francesco, Filley Eugene, Christopher Punita, Chen Fred K, Wuyyuru Varalakshmi, Sahel Jose, Stanga Paulo, Humayun Mark, Greenberg Robert J, Dagnelie Gislin, Argus II Study Group. The Argus II epiretinal prosthesis system allows letter and word reading and long-term function in patients with profound vision loss.. The British journal of ophthalmology. 2013. DOI: 10.1136/bjophthalmol-2012-301525.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.