DOI 10.1111/j.1540-8167.2008.01342.x

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2008

Effects of electrocautery on transvenous lead insulation materials.

Lim Kiam-Khiang, Reddy Shantanu, Desai Shrojal, Smelley Matthew, Kim Susan S, Beshai John F, Lin Albert C, Burke Martin C, Knight Bradley P

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Insulation defects are a leading cause of transvenous lead failure. The purpose of this study was to determine the effects of electrocautery on transvenous lead insulation materials.

METHODS

A preparation was done to simulate dissection of a transvenous lead from tissues. Radiofrequency energy was delivered using a standard cautery blade at outputs of 10, 20, and 30 W, for 3 and 6 seconds, using parallel and perpendicular blade orientations on leads with outermost insulations of silicone rubber, polyurethane, and silicone-polyurethane copolymer. Damage to each lead segment was classified after visual and microscopic analysis.

RESULTS

Significant insulation damage occurred to almost all polyurethane leads. Full insulation breaches were observed with 30 W regardless of application duration with a parallel direction and with all power outputs with a perpendicular direction. Thermal insulation damage to copolymer insulation was similar to that of the polyurethane leads. In contrast, there was no thermal damage to silicone leads, regardless of the power output and duration of power delivery. However, mechanical insulation damage was observed to all silicone leads when at least 20 W was applied in a direction perpendicular to the lead.

CONCLUSIONS

Polyurethane (PU55D) and copolymer materials have low thermal stability and are highly susceptible to thermal damage during cautery. Implanting physicians should be aware of the lead insulation materials being used during implant procedures and their properties. The use of direct contact cautery on transvenous leads should be minimized to avoid damage to the lead, especially on leads with polyurethane or copolymer outer insulations.

Для цитирования: Lim Kiam-Khiang, Reddy Shantanu, Desai Shrojal, Smelley Matthew, Kim Susan S, Beshai John F, Lin Albert C, Burke Martin C, Knight Bradley P. Effects of electrocautery on transvenous lead insulation materials.. Journal of cardiovascular electrophysiology. 2008. DOI: 10.1111/j.1540-8167.2008.01342.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.