Mechanical support of the left ventricle in ischemia induced left ventricular failure: an experimental study.
Аннотация
OBJECTIVE
this study compares the hemodynamic effects of intra-aortic balloon pumping (IABP), left ventricular assist device (LVAD), and extracorporeal membrane oxygenation (ECMO) in left ventricular failure in pigs.
METHODS
In 29 pigs weighing 12 +/- 0.7 kg left ventricular failure was induced by ligating the left anterior descending coronary artery. Eight animals served as controls. Eight pigs were treated by IABP, seven by LVAD, and six by ECMO. The study period lasted four hours. Hemodynamic and oxygen transport/uptake parameters were measured continuously or intermittently.
RESULTS
Six animals of the ECMO and LVAD groups survived the 4 hour period, but only 3 and 4 animals of the IABP and control groups survived (p less than 0.05). Cardiac index decreased about 48% and 22% in the control and IABP groups (p less than 0.05), whereas there was only a slight decrease in the ECMO (9%) and LVAD (14%) groups. Oxygen delivery fell significantly in the control and IABP groups (p less than 0.05), compared with only a slight change in the LVAD and ECMO groups.
CONCLUSION
ECMO is the most effective system for temporary circulatory support in severe ventricular failure. LVAD maintains cardiac output when pulmonary blood flow is provided. IABP is less efficient in supporting the failing heart, especially in the presence of severe ventricular arrhythmias.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.