DOI 10.1136/bmjophth-2022-001197
Improved screening of retinopathy of prematurity (ROP): development of a target product profile (TPP) for resource-limited settings.
Аннотация
BACKGROUND
As more preterm infants survive, complications of preterm birth, including retinopathy of prematurity (ROP), become more prevalent. ROP rates and blindness from ROP are higher in low-income and middle-income countries, where exposure to risk factors can be higher and where detection and treatment of ROP are under-resourced or non-existent. Access to low-cost imaging devices would improve remote screening capabilities for ROP.
METHODS
Target product profiles (TPPs) are developed early in the medical device development process to define the setting, target user and range of product requirements. A Delphi-like process, consisting of an online survey and consensus meeting, was used to develop a TPP for an ROP imaging device, collecting feedback on a proposed set of 64 product requirements.
RESULTS
Thirty-six stakeholders from 17 countries provided feedback: clinicians (72%), product developers (14%), technicians (6%) and other (8%). Thirty-six per cent reported not currently screening for ROP, with cited barriers including cost (44%), no training (17%) and poor image quality (16%). Among those screening (n=23), 48% use more than one device, with the most common being an indirect ophthalmoscope (87%), followed by RetCam (26%) and smartphone with image capture (26%). Consensus was reached on 53 (83%) product requirements. The 11 remaining were discussed at the consensus meeting, and all but two achieved consensus.
CONCLUSIONS
This TPP process was novel in that it successfully brought together diverse stakeholders to reach consensus on the product requirements for an ROP imaging devices. The resulting TPP provides a framework from which innovators can develop prototypes.