BACKGROUND

Due to potential thromboembolic complications, mechanical valves within the native heart are often considered contraindications to ventricular assist device (VAD) support.

METHODS

A retrospective review of VAD cases between June 1982 and March 1998 showed 8 patients with mechanical valves who were supported with Thoratec (Pleasanton, CA) VADs.

RESULTS

There were 6 males and 2 females ranging in age from 20 to 69 years (mean 49.8+/-5.6). Four patients were supported when they could not be weaned from cardiopulmonary bypass after reparative procedures and were thought to have reversible injuries. Four patients were supported as a bridge-to-cardiac transplantation. Two patients had mechanical mitral valves, 2 had aortic valve replacements, 1 had an aortic homograft and mechanical mitral valve, 2 had mechanical aortic and mitral prosthesis, and 1 patient had aortic, mitral, and tricuspid valves. The types of valvular prostheses were St. Jude (5 patients) and Bjork-Shiley (3 patients). Duration of support ranged from 3.0 to 150 days (mean 34 days). Four patients were supported with biventricular assist devices and 4 had left VADs. Dextran and intravenous heparin anticoagulation were used in the shorter duration patients, with warfarin being used in the bridge patients. One patient received warfarin and aspirin. At the time of autopsy or device removal, only 1 of the 12 mechanical intracardiac valves showed any evidence of thrombosis, including the aortic valves in 2 patients supported for 2 and 5 months. There were no clinical thromboembolic events. Four patients (50%) were discharged (1 weaned, 3 transplanted).

CONCLUSIONS

The 50% (4 of 8) survival rate compares favorably with the 44% (41 of 92) overall survival rate for our Thoratec patients (bridge plus recovery) who did not have mechanical prosthetic valves. These data suggest that patients with mechanical intracardiac valves can be supported for short durations with some additional risk, which is yet to be determined.

Для цитирования: Swartz M T, Lowdermilk G A, Moroney D A, McBride L R. Ventricular assist device support in patients with mechanical heart valves.. The Annals of thoracic surgery. 1999. DOI: 10.1016/s0003-4975(99)01110-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.