DOI 10.1177/11206721231210895

Медицинская библиотека · European journal of ophthalmology · 2023

Comparison between IOL MASTER 500 and MYAH with vector analysis in low and mild anterior corneal astigmatism.

Camellin Umberto, Franchina Francesco, Latino Gianluigi, Ninotta Ivan, Palino Paola, Meduri Alessandro, Aragona Pasquale

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: Evaluate the agreement between IOL Master 500 (Carl Zeiss Meditec AG, Germany) and MYAH (Topcon EU, Visia Imaging, Japan) in measuring axial length, keratometry, and anterior corneal astigmatism.

SETTING

Department of Biomedical and Dental Sciences and Morphofunctional Imaging, Ophthalmology Clinic, University of Messina, Messina, Italy.

METHODS

In this prospective study, 40 eyes (right eye 25, left eye 15) of 40 patients were included. Axial length (AL), keratometry (K1, K2), and anterior corneal astigmatism (ACA) were measured. AL, mean K (Kavg) and magnitude of ACA were compared using Bland - Altman plot analysis, parametric and nonparametric statistical analysis. The difference vector and angle of error of the ACA measured by the two devices were evaluated by vector analysis using polar diagrams.

RESULTS

Mean and standard deviation with IOL Master and with MYAH device was for AL 24.25 ± 1.22 mm and 24.25 ± 1.22 mm (p = .99), for Kavg 42.75 ± 1.53 D and 42.85 ± 1.52 D (p = .78), for Magnitude of ACA 1.00 ± 0.58 D and 0.89 ± 0.56 D (p = .38) respectively. High correlations were found for AL (R² = 0.999), Kavg (R² = 0.996), and ACA Magnitude (R² = 0.889). Bland-Altman analysis of the two devices found high agreement and absence of proportional bias (MYAH-IOL MASTER) were found between the two assessments for AL (bias = -0.0005 mm, p = .93), Kavg (bias = 0.0955 D, p = .76) and ACA (bias = 0.11 D, p = .41). Limit of agreement (upper/lower LoA, 95%CI) were respectively +0.057/-0.058 mm for AL, + 0.29/-0.09 D for Kavg and +0.49/-0.27 D for ACA. No statistical difference was found between the x-component and y-component of the ACA vector (p > .01), the difference vector (IOL MASTER-MYAH) was +0.14 D axis 159 with an absolute mean angle of error of 7.2 ± 7.5 degree.

CONCLUSIONS

The instruments appear to be interchangeable for measurements of AL, keratometry, and magnitude of ACA with high agreement between the two devices. Also, in the presence of low astigmatism, the two instruments give the same results in terms of ACA.

Для цитирования: Camellin Umberto, Franchina Francesco, Latino Gianluigi, Ninotta Ivan, Palino Paola, Meduri Alessandro, Aragona Pasquale. Comparison between IOL MASTER 500 and MYAH with vector analysis in low and mild anterior corneal astigmatism.. European journal of ophthalmology. 2023. DOI: 10.1177/11206721231210895.