DOI 10.1097/j.jcrs.0000000000001344
Comparison of toric intraocular lens alignment between femtosecond laser-assisted capsular marking and digital marking.
Аннотация
PURPOSE: To compare the accuracy of toric intraocular lens (IOL) alignment between femtosecond laser-assisted capsular marking and digital marking.
SETTING
Ruhr University Eye Clinic, Bochum, Germany.
DESIGN
Prospective clinical trial.
METHODS
In this study, 28 eyes of 23 patients, who underwent femtosecond laser-assisted cataract surgery with implantation of a toric IOL, were included. Intraoperatively, both femtosecond laser-assisted capsular marking and digital marking were applied simultaneously and compared in every case. The toric IOL was aligned to the capsular markings. Postoperatively, the axis of the capsular markings and toric IOL alignment was examined. Visual acuity and refractive outcomes were evaluated.
RESULTS
Both alignment methods were performed without intraoperative complications in all cases. 25 eyes were included in the final analysis. Misalignment was significantly lower with femtosecond laser-assisted capsular marking than with digital marking (1.71 ± 1.25 degrees vs 2.64 ± 1.70 degrees, P = .016). Deviation from the target axis of the toric IOL was 1.62 ± 1.24 degrees 4 to 6 weeks postoperatively. Postoperative uncorrected distance visual acuity was 0.14 ± 0.13 logMAR, and residual astigmatism was 0.3 ± 0.23 diopter (D) with an astigmatism ≤0.5 D in 93% of eyes.
CONCLUSIONS
Both methods showed excellent results for the alignment of toric IOLs. However, femtosecond laser-assisted capsular marking was significantly more precise than digital marking and showed good refractive results. In addition, capsular marking offers the possibility to avoid parallax error and evaluating postoperative IOL rotation.