Repeatability and Interobserver Reproducibility of a Swept-Source Optical Coherence Tomography for Measurements of Anterior, Posterior, and Total Corneal Power.

Lei Chak Seng, Lin Xuanqiao, Ning Rui, Yu Jinjin, Huang Xiaomin, Li Kexin, Wang Yiran, Savini Giacomo, Schiano-Lomoriello Domenico, Zhou Xingtao, Huang Jinhai

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The aim of this work is to evaluate the intraobserver repeatability and interobserver reproducibility of corneal power measurements obtained with a swept-source optical coherence tomographer (CASIA 2, Tomey, Japan) in healthy subjects.

METHODS

A total of 67 right eyes from 67 healthy subjects were enrolled. Two experienced observers measured each eye three times consecutively with the CASIA 2. Corneal power values were recorded as simulated keratometry, anterior, posterior, and total corneal power. Parameters were flattest keratometry (Kf), steepest keratometry (Ks), mean keratometry (Km), astigmatism magnitude, astigmatism power vectors J0 and J45. Intraobserver repeatability and interobserver reproducibility of the CASIA 2 were assessed by the within-subject standard deviation (Sw), test-retest repeatability (TRT), coefficients of variation (CoV), and intraclass correlation coefficients (ICCs). Double-angle plots were used for astigmatism vector analysis.

RESULTS

The CASIA 2 had high repeatability for all corneal power values, with Sw values ≤ 0.17 diopters (D), TRT ≤ 0.46 D, and ICCs ranging from 0.866 to 0.998. Interobserver reproducibility was also high, showing all Sw values ≤ 0.10 D, TRT ≤ 0.27 D, and ICCs ≥ 0.944. The reproducibility of the average of three consecutive measurements (Sw 0.01-0.10 D, TRT 0.03-0.27 D, ICC 0.944-0.998) was higher than the reproducibility of single measurements (Sw 0.01-0.17 D, TRT 0.03-0.47 D, ICC 0.867-0.996).

CONCLUSIONS

The CASIA 2 showed high intraobserver repeatability and interobserver reproducibility for anterior, posterior, and total corneal power measurements in 6.0-mm diameter area. In addition, we suggest that using the average of three consecutive measurements can improve reproducibility between observers, compared to single measurements only.

Для цитирования: Lei Chak Seng, Lin Xuanqiao, Ning Rui, Yu Jinjin, Huang Xiaomin, Li Kexin, Wang Yiran, Savini Giacomo, Schiano-Lomoriello Domenico, Zhou Xingtao, Huang Jinhai. Repeatability and Interobserver Reproducibility of a Swept-Source Optical Coherence Tomography for Measurements of Anterior, Posterior, and Total Corneal Power.. Ophthalmology and therapy. 2023. DOI: 10.1007/s40123-023-00815-9.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.