Comparison of a New Scheimpflug Camera and Swept-Source Optical Coherence Tomographer for Measurements of Anterior Segment Parameters.

Huang Xiaomin, Lin Xuanqiao, Yang Yizhou, Yu Jinjin, Wang Jiacheng, Li Kexin, Wang Yiran, Savini Giacomo, Schiano-Lomoriello Domenico, Zhou Xingtao, Huang Jinhai

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

This study evaluated the differences and agreement between a new Scheimpflug camera (Scansys) and a swept-source anterior segment optical coherence tomographer (CASIA 2) for measurements of the anterior segment of the eye in normal subjects.

METHODS

This prospective study included 84 eyes from 84 normal adult subjects who underwent three consecutive measurements with the Scansys and the CASIA 2 in random order. The mean keratometry (Km), astigmatism magnitude (AST), J0, and J45 vectors for both anterior and posterior corneal surfaces, central corneal thickness (CCT), thinnest corneal thickness (TCT), and anterior chamber depth (ACD) were obtained by both devices. The difference between these two devices was assessed using paired t test and violin plots. Bland-Altman plots and 95% limits of agreement (LoAs) were used to evaluate agreement.

RESULTS

No statistically significant differences between the two devices were found for the anterior AST, anterior J45, and posterior J45 (P > 0.05). The remaining parameters were statistically significant (P ≤ 0.05), but the differences not clinically significant. The violin plots showed that the distribution and probability density of the measured parameters were similar for both devices. Bland-Altman plots revealed high agreement for the measured parameters between the Scansys and CASIA 2, with narrow 95% LoAs.

CONCLUSIONS

In terms of assessing parameters for the anterior segment, our study indicated that Scansys and CASIA 2 generally showed significant agreement. The two devices used in this study's assessment of all the parameters can be used interchangeably in refractive analysis.

Для цитирования: Huang Xiaomin, Lin Xuanqiao, Yang Yizhou, Yu Jinjin, Wang Jiacheng, Li Kexin, Wang Yiran, Savini Giacomo, Schiano-Lomoriello Domenico, Zhou Xingtao, Huang Jinhai. Comparison of a New Scheimpflug Camera and Swept-Source Optical Coherence Tomographer for Measurements of Anterior Segment Parameters.. Ophthalmology and therapy. 2023. DOI: 10.1007/s40123-023-00813-x.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.