Reliability and Agreement of Keratometry Measurements Obtained With Eye Surface Profilometry and Partial Coherence Interferometry.

Martínez-Plaza Elena, Molina-Martín Ainhoa, Piñero David P

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this study was to assess the repeatability of keratometry parameters obtained using the Eye Surface Profiler (ESP) system and their agreement with the IOL Master 500 device.

METHODS

Seventy-one eyes of 71 healthy participants were evaluated. Three repeated measurements were performed using the ESP system. Simulated keratometry in the flat (SimKf) and steep (SimKs) meridians, astigmatism, and axis were obtained. The same parameters were measured using the IOL Master 500 device. The J0 and J45 vector components of the astigmatism were calculated. The intrasession repeatability was analyzed using within-subject SD (Sw) and intraclass correlation coefficient (ICC). Agreement was assessed using paired statistical tests and the Bland-Altman method.

RESULTS

The Sw was 0.07 mm, 0.04 mm, 0.51 D, 0.33 D, and 0.22 D, and the ICC was 0.96, 0.98, 0.74, 0.61, and 0.55 for SimKf, SimKs, astigmatism, J0, and J45, respectively. The mean difference and limits of agreement when comparing the ESP system with the IOL Master 500 device were 0.37 mm (0.08/0.66) for SimKf ( P < 0.001), 0.18 mm (0.00/0.35) for SimKs ( P < 0.001), -0.93 D (-2.42/0.56) for astigmatism ( P < 0.001), 0.51 D (-0.22/1.24) for J0 ( P < 0.001), and 0.06 D (-0.48/0.60) for J45 ( P = 0.09).

CONCLUSIONS

The ESP system provides consistent values for simulated keratometry, showing moderate consistency for astigmatism parameters. Contact lens practitioners should be aware that the ESP system and IOL Master 500 device provide different simulated keratometry from a clinically viewpoint.

Для цитирования: Martínez-Plaza Elena, Molina-Martín Ainhoa, Piñero David P. Reliability and Agreement of Keratometry Measurements Obtained With Eye Surface Profilometry and Partial Coherence Interferometry.. Cornea. 2023. DOI: 10.1097/ico.0000000000003348.