Cataract surgery causes biomechanical alterations to the eye detectable by Corvis ST tonometry.

Kato Yoshitake, Nakakura Shunsuke, Asaoka Ryo, Matsuya Kanae, Fujio Yuki, Kiuchi Yoshiaki, Ocular Biomechanics Study Group

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: Modern cataract surgery is generally considered to bring about modest and sustained intraocular pressure (IOP) reduction. However, the pathophysiological mechanism for this remains unclear. Moreover, a change in ocular biomechanical properties after surgery can affect the measurement of IOP. The aim of the study is to investigate ocular biomechanics, before and following cataract surgery, using Corvis ST tonometry (CST).

PATIENTS AND

METHODS

Fifty-nine eyes of 59 patients with cataract were analyzed. IOP with Goldmann applanation tonometry (IOP-G), axial length, corneal curvature and CST parameters were measured before cataract surgery and, up to 3 months, following surgery. Since CST parameters are closely related to IOP-G, linear modeling was carried out to investigate whether there was a change in CST measurements following cataract surgery, adjusted for a change in IOP-G.

RESULTS

IOP-G significantly decreased after surgery (mean±standard deviation: 11.8±3.1 mmHg) compared to pre-surgery (15.2±4.3 mmHg, P<0.001). Peak distance (the distance between the two surrounding peaks of the cornea at the highest concavity), maximum deformation amplitude (the movement of the corneal apex from the start of deformation to the highest concavity) and A1/A2 velocity (the corneal velocity during inward or outward movement) significantly increased after cataract surgery (P<0.05) while radius (the central curvature radius at the highest concavity) was significantly smaller following cataract surgery (P<0.05). Linear modeling supported many of these findings, suggesting that peak distance, maximum deformation amplitude and A2 velocity were increased, whereas A2 deformation amplitude and highest concavity time were decreased (after adjustment for IOP change), following cataract surgery.

CONCLUSION

Corneal biomechanical properties, as measured with CST, were observed to change significantly following cataract surgery.

TRIAL REGISTRATION: Japan Clinical Trials Registry UMIN000014370.

Для цитирования: Kato Yoshitake, Nakakura Shunsuke, Asaoka Ryo, Matsuya Kanae, Fujio Yuki, Kiuchi Yoshiaki, Ocular Biomechanics Study Group. Cataract surgery causes biomechanical alterations to the eye detectable by Corvis ST tonometry.. PloS one. 2017. DOI: 10.1371/journal.pone.0171941.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.