MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/1471-2474-11-139

Медицинская библиотека · BMC musculoskeletal disorders · 2010

Fixation strength of biocomposite wedge interference screw in ACL reconstruction: effect of screw length and tunnel/screw ratio. A controlled laboratory study.

Herrera Antonio, Martínez Fernando, Iglesias Daniel, Cegoñino José, Ibarz Elena, Gracia Luis

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Primary stability of the graft is essential in anterior cruciate ligament surgery. An optimal method of fixation should be easy to insert and provide great resistance against pull-out forces.A controlled laboratory study was designed to test the primary stability of ACL tendinous grafts in the tibial tunnel. The correlation between resistance to traction forces and the cross-section and length of the screw was studied.

METHODS

The tibial phase of ACL reconstruction was performed in forty porcine tibias using digital flexor tendons of the same animal. An 8 mm tunnel was drilled in each specimen and two looped tendons placed as graft. Specimens were divided in five groups according to the diameter and length of the screw used for fixation. Wedge interference screws were used. Longitudinal traction was applied to the graft with a Servohydraulic Fatigue System. Load and displacement were controlled and analyzed.

RESULTS

The mean loads to failure for each group were 295,44 N (Group 1; 9 x 23 screw), 564,05 N (Group 2; 9 x 28), 614,95 N (Group 3; 9 x 35), 651,14 N (Group 4; 10 x 28) and 664,99 (Group 5; 10 x 35). No slippage of the graft was observed in groups 3, 4 and 5. There were significant differences in the load to failure among groups (ANOVA/P < 0.001).

CONCLUSIONS

Longer and wider interference screws provide better fixation in tibial ACL graft fixation. Short screws (23 mm) do not achieve optimal fixation and should be implanted only with special requirements.

Для цитирования: Herrera Antonio, Martínez Fernando, Iglesias Daniel, Cegoñino José, Ibarz Elena, Gracia Luis. Fixation strength of biocomposite wedge interference screw in ACL reconstruction: effect of screw length and tunnel/screw ratio. A controlled laboratory study.. BMC musculoskeletal disorders. 2010. DOI: 10.1186/1471-2474-11-139.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.