Медицинская библиотека · Clinical chemistry · 1996

Automated HPLC screening of newborns for sickle cell anemia and other hemoglobinopathies.

Eastman J W, Wong R, Liao C L, Morales D R

СтатьяПеревод пока не подготовлен

Аннотация

Automated HPLC is used to test dried blood-spot specimens from newborns for hemoglobins (Hb) F, A, S, C, E, and D. We present the method and report on its performance determined during >4 years of testing 2.5 x 10(6) newborns. The method features automated derivation of presumptive phenotypes; quantitative quality control and proficiency testing; throughput of one specimen per minute; small sample volume; hemoglobin concentrations quantified with an interlaboratory CV of 14-18%; retention times with interlaboratory CV of <2% and matching, within +/- 0.03 min, of laboratories and reagent lots; control of peak resolution; 0.5% detection limit for Hb S and C, and 1.0% for Hb F, A, E, and D; few interferences; and negligible background and carryover. Shortcomings of the method are the absence of microplate barcode identification and the need for manually pipetting the sample eluate into the microplate.

Для цитирования: Eastman J W, Wong R, Liao C L, Morales D R. Automated HPLC screening of newborns for sickle cell anemia and other hemoglobinopathies.. Clinical chemistry. 1996.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.