Optimization of the skin multiple analyte profile bioanalytical method for determination of skin biomarkers from D-Squame tape samples.

Hendrix S W, Miller K H, Youket T E, Adam R, O'Connor R J, Morel J G, Tepper B E

Клиническое исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND/PURPOSE: This work was performed to optimize extraction conditions for D-Squame tape skin samples for use in the skin multiple analyte profile (SkinMAP) method, a Linco Research Corporation bead-based assay for skin analytes. The experiments were designed to help identify sources of variability during extraction that may be amenable to further control.

METHODS

Two experimental designs were used to study factors influencing the extraction of skin samples from D-Squame tapes. Visually healthy skin samples were obtained from both female and male adult volar forearms. Factors studied in two experiments included: four surfactant (SDS) levels (0.02-0.2%), two buffer types [Citrate-phosphate buffered saline at pH 5.5, phosphate-buffered saline (PBS) at pH 7.4], two buffer volumes (1.0, 1.5 mL), two propylene glycol (PG) levels (0.1%, 1.0%), two extraction temperatures (7-10 degrees C, 22-30+ degrees C), two extraction times (30, 60 min), and location in sonication bath (two vectors). The response biomarkers were cortisol, fibronectin, human serum albumin, involucrin, keratin-6 and keratins 1, 10. Skin sampling sites were also evaluated as sources of variation.

RESULTS

There was no single set of extraction conditions in our experiments that maximized recovery of all the biomarkers. SDS level had the most consistently significant (P<0.05) and directional effects on biomarker recoveries. In general, higher SDS resulted in higher recovery of all biomarkers. There was less consistency and fewer significant results for the other extraction factors.

CONCLUSIONS

These data enable us to better manage SkinMAP studies and interpret their results. The use of 1.5 mL PBS containing 0.2% SDS and 0.5% PG with 30 min sonication at low (near 4 degrees C) temperature is optimal for the quantitation of a range of SkinMAP analytes. In order to protect researchers from obtaining inflated false positive rates, it is crucial to design such studies and analyze the data using appropriate statistical methodology, especially for those studies involving only a small number of subjects.

Для цитирования: Hendrix S W, Miller K H, Youket T E, Adam R, O'Connor R J, Morel J G, Tepper B E. Optimization of the skin multiple analyte profile bioanalytical method for determination of skin biomarkers from D-Squame tape samples.. Skin research and technology : official journal of International Society for Bioengineering and the Skin (ISBS) [and] International Society for Digital Imaging of Skin (ISDIS) [and] International Society for Skin Imaging (ISSI). 2007. DOI: 10.1111/j.1600-0846.2007.00235.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.