Prediction of occluded-side M1 vessel diameter using pre-procedural magnetic resonance angiography.

Kitano Takaaki, Ikeda Hiroyuki, Hata Hidenobu, Akaike Natsuki, Tanimura Mai, Yokochi Yasunori, Osuki Takuya, Kaneko Ryosuke, Uezato Minami, Kinosada Masanori, Kurosaki Yoshitaka, Chin Masaki

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To determine whether the occluded-side M1 vessel diameter on post-procedural digital subtraction angiography (DSA) can be predicted from that on pre-procedural magnetic resonance angiography (MRA) and its usefulness for aspiration catheter selection.

METHODS

We retrospectively analyzed 41 cases of M1 occlusion undergoing mechanical thrombectomy with successful recanalization and pre- and post-procedural MRA at our institution between January 2020 and December 2024. The relationship between the catheter-to-vessel ratio (CVR), calculated from aspiration catheter's outer diameter and occluded-side M1 vessel diameter on pre-procedural MRA, and first-pass effect was assessed.

RESULTS

The mean occluded-side M1 vessel diameter was 2.2 ± 0.3 mm on pre-procedural MRA and 2.4 ± 0.4 mm on post-procedural DSA; the latter showed a strong correlation with the former (r = 0.925, p < 0.001), and the regression equation was as follows: occluded-side M1 vessel diameter on post-procedural DSA (mm) = 0.19 + 1.025 × occluded-side M1 vessel diameter on pre-procedural MRA (mm) (R² = 0.856). The occluded-side M1 vessel diameter on post-procedural DSA approximated from the mean occluded-side M1 vessel diameter on pre-procedural MRA was pre-procedural MRA measurement + 0.2 mm. First-pass effect was achieved in 66.7 % of cases with CVR ≥ 0.80 and in 25.0 % of cases with CVR < 0.80 (p = 0.049).

CONCLUSION

When treating M1 occlusion, adding 0.2 mm to the occluded-side M1 vessel diameter on pre-procedural MRA measurement allows prediction of the actual M1 vessel diameter. Selecting an aspiration catheter with CVR ≥ 0.80 may help achieve first-pass effect.

Для цитирования: Kitano Takaaki, Ikeda Hiroyuki, Hata Hidenobu, Akaike Natsuki, Tanimura Mai, Yokochi Yasunori, Osuki Takuya, Kaneko Ryosuke, Uezato Minami, Kinosada Masanori, Kurosaki Yoshitaka, Chin Masaki. Prediction of occluded-side M1 vessel diameter using pre-procedural magnetic resonance angiography.. Clinical neurology and neurosurgery. 2026. DOI: 10.1016/j.clineuro.2026.109345.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.