MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11325-017-1583-5

Медицинская библиотека · Sleep & breathing = Schlaf & Atmung · 2017

Ambulatory screening for obstructive sleep apnea in patients with resistant arterial hypertension.

Kamasová Monika, Václavík Jan, Václavík Tomáš, Hobzová Milada, Kociánová Eva, Táborský Miloš

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Obstructive sleep apnea (OSA) is one of the most common causes of secondary arterial hypertension. It is important to rule out OSA as a cause of resistant hypertension. The ApneaLink device is a simple and cost-efficient outpatient examination, but its usefulness in screening OSA in resistant hypertension has not yet been evaluated.

METHODS

A total of 69 patients with resistant arterial hypertension were enrolled. Patients underwent a physical examination, including the use of ApneaLink, followed by respiratory polygraphy. The presence of OSA was assessed by the apnea-hypopnea index (AHI), oxygen desaturation index (ODI), mean nocturnal desaturation (SpO2), and percentage of sleep time with SpO2 less than 90%.

RESULTS

There was no significant difference between the values of AHI found during the use of ApneaLink and respiratory polygraphy (mean 30.4 ± 21.7 vs. 37.2 ± 20.9, P = 0.07). ApneaLink had 77.3% sensitivity and 100% specificity to diagnose OSA with the area under the ROC curve 0.866 (P < 0.001). We also found no significant difference in mean SpO2 (91.3 ± 2.5 vs. 90.9 ± 3.3%, P = 0.22). The ODI evaluated via ApneaLink was significantly lower than by the polygraphy (31.1 ± 18.3 vs. 43.9 ± 24.8, P < 0.001), while the measured percentage of sleep time with SpO2 less than 90% was higher (31.8 ± 23.7 vs. 23.3 ± 24.4, P = 0.001). The severity of OSA was correctly determined by ApneaLink in 50.7% of patients, underestimated in 23.2% and overestimated in 26.1%.

CONCLUSIONS

The use of ApneaLink is a suitable method for screening the presence of OSA in patients with resistant hypertension, but to accurately assess the severity of OSA, respiratory polygraphy or polysomnography is required.

Для цитирования: Kamasová Monika, Václavík Jan, Václavík Tomáš, Hobzová Milada, Kociánová Eva, Táborský Miloš. Ambulatory screening for obstructive sleep apnea in patients with resistant arterial hypertension.. Sleep & breathing = Schlaf & Atmung. 2017. DOI: 10.1007/s11325-017-1583-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.