Transcatheter heart valve failure: a systematic review.

Mylotte Darren, Andalib Ali, Thériault-Lauzier Pascal, Dorfmeister Magdalena, Girgis Mina, Alharbi Waleed, Chetrit Michael, Galatas Christos, Mamane Samuel, Sebag Igal, Buithieu Jean, Bilodeau Luc, de Varennes Benoit, Lachapelle Kevin, Lange Ruediger, Martucci Giuseppe, Virmani Renu, Piazza Nicolo

СтатьяПоддержка исследования: вне правительства СШАСистематический обзорПеревод пока не подготовлен

Аннотация

AIMS: A comprehensive description of transcatheter heart valve (THV) failure has not been performed. We undertook a systematic review to investigate the aetiology, diagnosis, management, and outcomes of THV failure.

METHODS AND

RESULTS

The systematic review was performed in accordance with the PRISMA guidelines using EMBASE, MEDLINE, and Scopus. Between December 2002 and March 2014, 70 publications reported 87 individual cases of transcatheter aortic valve implantation (TAVI) failure. Similar to surgical bioprosthetic heart valve failure, we observed cases of prosthetic valve endocarditis (PVE) (n = 34), structural valve failure (n = 13), and THV thrombosis (n = 15). The microbiological profile of THV PVE was similar to surgical PVE, though one-quarter had satellite mitral valve endocarditis, and surgical intervention was required in 40% (75% survival). Structural valve failure occurred most frequently due to leaflet calcification and was predominantly treated by redo-THV (60%). Transcatheter heart valve thrombosis occurred at a mean 9 ± 7 months post-implantation and was successfully treated by prolonged anticoagulation in three-quarters of cases. Two novel causes of THV failure were identified: late THV embolization (n = 18); and THV compression (n = 7) following cardiopulmonary resuscitation (CPR). These failure modes have not been reported in the surgical literature. Potential risk factors for late THV embolization include low prosthesis implantation, THV undersizing/underexpansion, bicuspid, and non-calcified anatomy. Transcatheter heart valve embolization mandated surgery in 80% of patients. Transcatheter heart valve compression was noted at post-mortem in most cases.

CONCLUSION

Transcatheter heart valves are susceptible to failure modes typical to those of surgical bioprostheses and unique to their specific design. Transcatheter heart valve compression and late embolization represent complications previously unreported in the surgical literature.

Для цитирования: Mylotte Darren, Andalib Ali, Thériault-Lauzier Pascal, Dorfmeister Magdalena, Girgis Mina, Alharbi Waleed, Chetrit Michael, Galatas Christos, Mamane Samuel, Sebag Igal, Buithieu Jean, Bilodeau Luc, de Varennes Benoit, Lachapelle Kevin, Lange Ruediger, Martucci Giuseppe, Virmani Renu, Piazza Nicolo. Transcatheter heart valve failure: a systematic review.. European heart journal. 2014. DOI: 10.1093/eurheartj/ehu388.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.