DOI 10.1097/00006324-200306000-00009

Медицинская библиотека · Optometry and vision science : official publication of the American Academy of Optometry · 2003

Measuring central corneal thickness with ultrasound pachymetry.

Realini Tony, Lovelace Kim

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine how many ultrasound pachymetry measurements are required to obtain a reasonable estimate of central corneal thickness.

METHODS

Five central corneal thickness measurements were obtained from each eye of 101 patients with normal corneas. The means of fewer than five readings were compared with the mean of five values for each eye to determine the incremental clinical value of additional measurements.

RESULTS

Compared with the mean of five values, fewer than three measurements per eye provided a mean central corneal thickness value with inadequate confidence. Conversely, means of more than three readings provided little additional confidence over means of three values.

CONCLUSIONS

Central corneal thickness by ultrasound pachymetry can be adequately assessed in the majority of eyes by taking three measurements per eye.

Для цитирования: Realini Tony, Lovelace Kim. Measuring central corneal thickness with ultrasound pachymetry.. Optometry and vision science : official publication of the American Academy of Optometry. 2003. DOI: 10.1097/00006324-200306000-00009.